Agilent / Keysight 85039A Calibration Kit, Type-F
Agilent / Keysight RF Calibration Equipment. Frequency Range: DC to 3 GHz.
- Model
- Agilent / Keysight 85039A
- Calibration
- Available on request
- Price
- $1,260
- Status
- In stock
- RFQ ref
- 85039A
This unit
The Agilent/Keysight 85039A is a Type-F calibration kit engineered for precision calibration of 75-ohm RF and microwave test equipment. It delivers traceable calibration standards and adapters across DC to 3 GHz, enabling accurate S-parameter measurements and network analyzer validation on systems equipped with Type-F interfaces.
Technical Specifications
- Frequency Range: DC to 3 GHz
- Impedance: 75 ohms
- Connector Type: Type-F male standards
- Operating Temperature: +15 to +35°C
- Storage Temperature: −40 to +75°C
Key Features The kit comprises three primary calibration standards. The Type-F Male Open standard exhibits phase deviation of ±0.55° from DC to 1 GHz and ±1.30° from 1 GHz to 3 GHz. The Type-F Male Short standard provides phase deviation of ±0.60° from DC to 1 GHz and ±1.00° from 1 GHz to 3 GHz. The Type-F Male Load delivers return loss exceeding 32 dB (ρ ≤ 0.025) from 1.3 GHz to 3 GHz, with return loss ≥ 45 dB (ρ ≤ 0.006) from DC to 1 GHz and ≥ 38 dB (ρ ≤ 0.013) from 1 GHz to 3 GHz. Included adapters extend system flexibility. Type-N to Type-F adapters provide return loss ≥ 38 dB (ρ ≤ 0.013) from DC to 1 GHz and > 26 dB (ρ < 0.050) from 1.3 GHz to 3 GHz. Type-F to Type-F female adapters specify return loss ≥ 40 dB (ρ ≤ 0.013) from DC to 1 GHz and ≥ 32 dB (ρ ≤ 0.025) from 1 GHz to 3 GHz.
Typical Applications Calibration of network analyzer systems and 75-ohm test equipment. S-parameter measurement validation across RF and microwave frequency bands.
Compatibility & Integration Designed for Keysight 8752C, 8753 Series (8753A, 8753B, 8753C, 8753D), 75-ohm test sets, and Agilent 8712E Network Analyzer Systems. Connectors comply with SCTE specifications; center conductor dimensions range from 0.56 to 1.07 mm (0.022 to 0.042 inches). ESD precautions required during handling.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | DC to 3 GHz |
Agilent / Keysight 85039A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 85039A. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Agilent / Keysight 85039A price
This Agilent / Keysight 85039A is listed at $1,259.8. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight 85039A calibration cost
NIST-traceable calibration for the Agilent / Keysight 85039A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →