Agilent/Keysight N2782B
Agilent / Keysight Laboratory Measurement Instruments. Accuracy: ±1.0% of reading ±10 mA, Sensitivity: 0.1 V/A (10:1 attenuation).
- Model
- Agilent / Keysight N2782B
- Calibration
- Available on request
- Price
- $2,250
- Status
- In stock
- RFQ ref
- N2782B
This unit
The Agilent/Keysight N2782B is a 50 MHz AC/DC current probe that delivers non-invasive, accurate current measurements from DC to 50 MHz using hybrid Hall-effect and AC current transformer technology. It measures continuous RMS currents up to 30 Arms with pulse capability to 50 A, while maintaining ±1.0% amplitude accuracy and low insertion impedance. The probe connects directly to oscilloscope 1 MΩ inputs via a 1.5 m BNC cable and includes a demagnetize function for core de-magnetization and zero adjustment for offset correction.
Technical Specifications
- Measurement Range: DC to 50 MHz bandwidth; 5 mA minimum measurable current
- Current Capacity: 30 Arms continuous; 50 A maximum pulse current
- Amplitude Accuracy: ±1.0% of reading ±10 mA
- Rise Time: 7 ns or less
- Output Sensitivity: 0.1 V/A (10:1 attenuation)
- Output Impedance: 1 MΩ when terminated into 1 MΩ oscilloscope input
- Safety Rating: 300 V CAT I, 300 V CAT III
- Maximum Conductor Diameter: 5 mm
- Cable Length: 1.5 m probe to BNC
Key Features
- Hybrid technology combining Hall-effect sensor with AC current transformer for DC and AC measurement
- High signal-to-noise ratio for low-level milliampere measurements
- Demagnetize function removes residual core magnetism
- Zero adjustment control corrects voltage offset and temperature drift
- Split-core design enables non-invasive clip-on/clip-off connection
- Overload protection safeguards probe integrity
- Compact dimensions: 22.9 cm × 9.7 cm × 2.5 cm; 0.23 kg weight
Typical Applications Motor drives, switching power supplies, inverters, controllers, disk drives, LCD displays, electronic ballasts, and amplifiers driving inductive loads. Ideal for steady-state and transient current analysis in power electronics.
Compatibility & Integration Direct connection to Keysight 1000, 3000, InfiniiVision 2000X/3000X/4000X/5000/6000/7000 Series, and Infiniium 8000/9000 Series oscilloscopes. For Infiniium 90000A Series, requires E2697A adapter; for 90000 X-Series, requires N5449A adapter. If oscilloscope input is limited to 50 Ω, use E2697A adapter for 50 Ω to 1 MΩ conversion. Requires Keysight N2779A external power supply; one supply powers up to three N278xB probes.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | DC to 50 MHz | Voltage / Current Range: 5 mA to 30 A RMS, 50 A peak |
Agilent / Keysight N2782B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight N2782B. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Agilent / Keysight N2782B price
This Agilent / Keysight N2782B is listed at $2,250. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight N2782B calibration cost
NIST-traceable calibration for the Agilent / Keysight N2782B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →