Stanford Research PS370 – High Voltage Power Supply
Stanford Research High Voltage Power Supplies. Current: 500 µA, Accuracy: 0.01% + 0.05% of full scale or 0.06% of full scale, Accuracy: ±1 V typical.
- Model
- Stanford Research PS370
- Calibration
- Available on request
- Price
- $1,350
- Status
- In stock
- RFQ ref
- PS370
This unit
The Stanford Research Systems PS370 is a high-voltage DC power supply engineered for precision laboratory and research applications requiring stable, adjustable negative voltage outputs. Operating across −100 V to −20 kV with maximum current of 500 µA and 10 W power output, the PS370 delivers exceptional voltage stability of 0.01% per hour with output ripple below 0.01% of full scale. The unit combines arc and short-circuit protection with programmable voltage, current limit, and current trip settings, making it suitable for sensitive experiments demanding low-noise performance.
Technical Specifications
Voltage Output- Output range: −100 V to −20 kV
- Maximum current: 500 µA
- Power output: 10 W
- Voltage set accuracy: 0.01% + 0.05% of full scale or 0.06% of full scale
- Voltage display accuracy: ±1 V typical, ±2 V maximum
- Voltage resolution and resetability: 1 V
- Voltage regulation: 0.001% for ±10% line change; 0.04% for 100% load change
- Output ripple (rms): <0.01% of full scale (300 Hz to 300 kHz)
- Slew rate: 14,000 V/s typical
- Discharge time: <6 seconds to <1% of full-scale voltage, no load
- Stored charge: <40 µC maximum Current Control
- Current limit range: 0 to 105% of full scale
- Current trip range: 10 µA to 105% of full scale
- Trip response time: <10 ms (excluding stored output charge)
- Current set accuracy: 1 µA or 0.5% of full scale
- Current resolution: 1 µA
- Current display accuracy: ±1 µA typical, ±2 µA maximum Stability & Performance
- Stability: 0.01% per hour, <0.03% per 8 hours
- Temperature drift: 50 ppm/°C from 10 to 40 °C typical
- Recovery time: 12 ms for 40% step load change
- Warm-up period: 1 hour for all specifications to apply
Key Features
- Dual large LED displays for voltage and current monitoring
- GPIB (IEEE-488) and RS-232 communication interfaces
- Analog I/O: 0 to +10 VDC input for programming; dual BNC outputs for voltage and current monitoring (0 to +10 V, 10 mA drive, 1 Ω output impedance)
- Front-panel indicators for limit state, remote mode, and rear analog input status
- Programmable protection with voltage limit, current limit, and current trip functions
Typical Applications
- High-voltage research and laboratory experiments
- Sensitive instrumentation requiring low ripple and noise
- Applications demanding precise voltage regulation and current control
Compatibility & Integration The PS370 integrates with standard laboratory automation systems via GPIB and RS-232 interfaces. Analog programming and monitoring via BNC connectors enable integration with external control systems and data acquisition equipment.
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | −100 V to −20 kV, 0 to 500 µA |
Stanford Research PS370 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Stanford Research PS370. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Stanford Research PS370 price
This Stanford Research PS370 is listed at $1,350. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Stanford Research PS370 calibration cost
NIST-traceable calibration for the Stanford Research PS370 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →