Ando AQ6330 1200-1700nm Optical Spectrum Analyzer
Ando AQ6330 1200-1700nm Optical Spectrum Analyzer,Compact and efficient optical spectrum analyzer for 1200 to 1700nm range. Offers high dynamic range and fast response time. Ideal for analyzing DWDM systems, laser sources, and optical filters in R&D and manufacturing.
- Model
- Ando AQ6330
- Calibration
- Available on request
- Price
- $9,780
- Status
- In stock
- RFQ ref
- AQ6330
This unit
The Ando AQ6330 is a portable optical spectrum analyzer for field installation and maintenance of WDM optical transmission systems. Operating across 1200–1700 nm, it delivers high-resolution spectral analysis in a compact 8 kg form factor. The instrument combines wavelength accuracy to ±0.05 nm (1500–1570 nm range) with ±0.3 dB level accuracy and supports measurements up to +20 dBm.
Built-in wavelength calibration and advanced WDM analysis of up to 100 channels eliminate dependence on external calibration sources, while a 7.8-inch color LCD and long-term monitoring capability enable detailed signal characterization at remote sites.
Technical Specifications
Wavelength Performance- Range: 1200–1700 nm
- Accuracy: ±0.05 nm (1500–1570 nm); ±0.02 nm linearity (1550 nm range); ±0.3 nm (full range)
- Level accuracy: ±0.3 dB at 1310/1550 nm
- Polarization dependency: ±0.05 dB Dynamic Range & Sensitivity
- Stray-light dynamic range: 40 dB (1523 nm, ±1.0 nm peak, 0.08 nm resolution); 30 dB (1523 nm, ±0.4 nm peak, 0.08 nm resolution)
- Return loss: 30 dB or better at 1310/1550 nm
- Maximum power measurement: +20 dBm (100 mW) WDM & Monitoring
- Multi-channel analysis: up to 100 WDM channels with wavelength, level, and SNR measurement
- Long-term monitoring: wavelength, level, and SNR tracking per channel Display & Storage
- 7.8-inch color LCD (640 × 480 pixels)
- Internal memory: 2 MB
- External storage: 3.5-inch floppy drive (1.44 MB)
Key Features
- Built-in wavelength standard for calibration without external source
- High-speed sweep for rapid measurement acquisition
- Single-mode fiber (10/125 µm) compatible with FC-PC input connector
- RS-232C and GP-IB remote control interfaces
- PCMCIA expansion slots (Type 3 × 1 or Type 2 × 2)
Typical Applications
- WDM system installation and validation
- Optical transmission line commissioning and troubleshooting
- Field-based spectral monitoring and long-term channel tracking
- Low-power signal detection in deployed networks
Compatibility & Integration Power: 100–120 V / 200–240 V AC, 50/60 Hz (~100 VA). Dimensions: 300 mm W × 200 mm H × 220 mm D. Operating temperature: 0–50°C. Keyboard (IBM), mouse (PS/2), VGA video, and Centronix printer ports enable laboratory and field integration.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '+20 dBm maximum | Voltage / Current Range: ~100 VA, 100–120 V / 200–240 V AC, 50/60 Hz | Wavelength Range: 1200 nm to 1700 nm |
Frequently Asked Questions
- What does the Ando AQ6330 measure and what wavelength range does it cover?
- The Ando AQ6330 is a compact optical spectrum analyzer designed to measure optical power across the 1200–1700 nm infrared band, covering the critical S-, C-, and L-bands used in DWDM telecommunications systems. It delivers high dynamic range performance with a maximum power measurement capability of +20 dBm, making it ideal for characterizing laser sources, optical filters, and WDM channel performance in both R&D and manufacturing environments. The instrument's fast response time supports efficient production testing and real-time system diagnostics.
- What condition is this used AQ6330 in and what has been verified?
- This unit is offered in used condition and has been received, inventoried, and visually inspected by our team at Aumictech. We verify basic mechanical integrity and functionality before listing; however, detailed performance validation against Ando's original specifications is available through our optional calibration service. As with all used test equipment, we recommend calibration prior to critical measurement applications to ensure traceability and accuracy.
- Is calibration available for the AQ6330 and what does it cost?
- Yes, we offer calibration services for the AQ6330 starting at $450, available as a separate add-on to your purchase. Our calibration covers the wavelength accuracy and power measurement response across the 1200–1700 nm range to verify performance against manufacturer specifications. Exact pricing and turnaround time depend on the scope of calibration selected; contact our sales team for a detailed quote and estimated lead time for your specific requirements.
- What interfaces and connectivity does the AQ6330 have for remote operation?
- The Ando AQ6330 features GPIB (IEEE-488) connectivity for integration into automated test systems and data acquisition setups common in optical communications labs. The instrument operates from standard AC mains (100–120 V or 200–240 V, 50/60 Hz) with typical power consumption around 100 VA, allowing flexible bench deployment. For specific software compatibility, driver availability, and any legacy interface considerations, we recommend consulting the original user manual or contacting Aumictech with your system integration details.
Ando AQ6330 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Ando AQ6330. Calibration covers wavelength accuracy at 850, 1310, 1550, and 1625 nm, power level accuracy, resolution bandwidth (RBW) accuracy, and dynamic range verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Ando AQ6330 faults: wavelength motor failure, internal light source dimming, detector sensitivity loss. Common repairs include wavelength motor servicing, internal broadband source replacement, and detector recalibration after extended use.
Ando AQ6330 price
This Ando AQ6330 is listed at $9,780. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Ando AQ6330 calibration cost
NIST-traceable calibration for the Ando AQ6330 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers wavelength accuracy at 850, 1310, 1550, and 1625 nm, power level accuracy, resolution bandwidth (RBW) accuracy, and dynamic range verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →