Agilent / Keysight E3646A DC Power Supply

The Agilent/Keysight E3646A is a reliable DC power supply providing stable and precise voltage and current for a variety of electronic testing and development applications. It features dual outputs, low noise, and excellent regulation.

Model
Agilent / Keysight E3646A
Calibration
Available on request
Price
$1,500
Status
In stock
RFQ ref
E3646A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Keysight E3646A is a 60 W dual-output, dual-range programmable DC power supply engineered for R&D, design verification, and production testing. Each output independently selects between 0–8 Vdc @ 0–3 A or 0–20 Vdc @ 0–1.5 A, with voltage programming accuracy of ±0.05% + 10 mV and load regulation under 0.01% + 3 mV. Output voltage ripple remains below 0.5 mVrms / 5 mVpp, while transient response recovers within 50 µsec following load step changes.


Readback accuracy for voltage is ±0.05% + 2 counts; current readback is ±0.15% + 5 mA on Output 1 and ±0.15% + 10 mA on Output 2.


Technical Specifications



  • Output Power: 60 W total

  • Dual-Range Outputs: 0–8 Vdc @ 0–3 A or 0–20 Vdc @ 0–1.5 A per output

  • Voltage Programming Accuracy: ±0.05% + 10 mV

  • Load Regulation (Voltage): < 0.01% + 3 mV

  • Load Regulation (Current): < 0.01% + 250 µA

  • Voltage Ripple & Noise: < 0.5 mVrms / < 5 mVpp

  • Current Noise: < 4 mArms

  • Transient Response: < 50 µsec (full to half load, within 15 mV)

  • Voltage Readback Accuracy: ±0.05% + 2 counts

  • Current Readback Accuracy: ±0.15% + 5 mA (Output 1); ±0.15% + 10 mA (Output 2)

  • Stability (8 hours, constant load/temperature): Voltage < 0.02% + 2 mV; Current < 0.02% + 3 mA

  • Over-Voltage Protection Accuracy: ±(0.5% + 0.5 V)

  • OVP Activation Time: < 1.5 msec (≥ 3 V); < 10 msec (< 3 V)

  • AC Input: 115 Vac ±10%, 47–63 Hz (standard); options for 100 Vac or 230 Vac

  • Maximum Input Power: 210 VA

  • Common Mode Current: < 1.5 µArms


Key Features



  • Front and rear output terminals for flexible test-point connectivity

  • Remote sensing capability with 1 V maximum load lead voltage drop

  • Vacuum fluorescent display with front-panel rotary knob and keypads

  • Dual-range selectable outputs on each channel

  • Programmable over-voltage protection with adjustable trip threshold

  • Fan-cooled with automatic speed control for reduced noise

  • Safety Class I with protective earth terminal

  • User-replaceable power line fuse


Interfaces & Control



  • GPIB (IEEE-488) - standard

  • RS-232 - standard

  • SCPI (Standard Commands for Programmable Instruments) compatible


Typical Applications



  • Component and subsystem design verification

  • Automated production test systems

  • Research and development bench work

  • Sensitivity and margin testing of electronic circuits

Agilent / Keysight E3646A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight E3646A. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Agilent / Keysight E3646A price

This Agilent / Keysight E3646A is listed at $1,500. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / Keysight E3646A calibration cost

NIST-traceable calibration for the Agilent / Keysight E3646A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents