Anritsu MU909015C – µOTDR Module, 3-Wavelength, 35 dB class OTDR For MT9090A Network Master Series Mainframes
The Anritsu MU909015C is a µOTDR module designed for use with MT9090A Network Master series mainframes. This 3-wavelength OTDR module offers a dynamic range of 35 dB, making it suitable for a variety of optical fiber testing applications. It is ideal for characterizing and troubleshooting optical networks.
- Model
- Anritsu MU909015C
- Calibration
- Available on request
- Price
- $3,800
- Status
- In stock
- RFQ ref
- MU909015C
This unit
The Anritsu MU909015C is a 3-wavelength µOTDR module engineered for the MT9090A Network Master series mainframes. This compact handheld instrument delivers up to 38 dB dynamic range and characterizes single-mode optical fiber from 0.5 km to 250 km, supporting installation, maintenance, and troubleshooting across premise, access, metro, core, and PON-based FTTx networks up to 1×64 split ratios.
Technical Specifications
Optical Wavelengths: 1310 nm, 1490 nm, 1550 nm, 1625 nm, 1650 nm
Dynamic Range: Up to 38 dB (20 µs pulse width); 35 dB class; 25 dB to 38 dB
Distance Range: 0.5 km to 250 km
Pulse Width Options: 5 ns, 10 ns, 20 ns, 50 ns, 100 ns, 500 ns, 1 µs, 2 µs, 5 µs, 10 µs, 20 µs
Sampling Resolution: 2 cm
Sampling Points: Less than 250,001 points
Dead Zone (Fresnel): ≤ 0.8 m typical
Dead Zone (Backscatter): ≤ 4.0 m at 1310 nm; ≤ 4.5 m at 1490/1550/1625/1650 nm (typical)
Distance Accuracy: ± 1 m ± 3 × measurement distance × 10⁻⁵ ± marker resolution
Wavelength Accuracy: ± 20 nm & ± 15 nm
Linearity: ± 0.05 dB or ± 0.1 dB
Return Loss Measurement: 14.8 ± 2 dB
IOR Setting Range: 1.3000 to 1.7000
Measurement Update Rate: 0.25 seconds real-time
Trace Format: Universal Telcordia SR-4731 (.SOR) issue 2
Key Features
Interchangeable FC and SC/UPC connectors
Pass/fail analysis with automatic threshold-based results
Multiple measurement modes: loss (2-point, splice loss, dB/km), LSA, ORL, event
4.3-inch TFT color LCD display, 480 × 272 pixels resolution
Bluetooth, Wi-Fi, Ethernet, and USB interfaces
Web browser remote control via USB-Ethernet converter
Fiber Type: Single Mode 10 µm/125 µm SMF (ITU-T G.652)
Typical Applications
Fiber installation and commissioning
Network maintenance and fault location
PON deployment and split-ratio characterization
Metro and core network troubleshooting
Compatibility & Integration The MU909015C integrates with MT9090A mainframes and operates on AC 100–240 V (50/60 Hz) or DC 9 V. NiMH battery options enable field deployment. Module weight: approximately 700 g; dimensions: 190 mm × 96 mm × 48 mm.
Anritsu MU909015C — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu MU909015C. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu MU909015C faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Anritsu MU909015C price
This Anritsu MU909015C is listed at $3,800. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu MU909015C calibration cost
NIST-traceable calibration for the Anritsu MU909015C is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →