HP / Agilent 81632B Power Sensor Module (single channel)

The HP/Agilent 81632B is a single-channel Power Sensor Module designed for optical power measurements. It is compatible with HP/Agilent lightwave measurement systems, offering accurate and reliable power readings in various optical communication applications. This module provides a wide dynamic range and supports different wavelength ranges, making it suitable for testing optical components and systems.

Model
HP / Agilent 81632B
Calibration
Available on request
Price
$1,260
Status
In stock
RFQ ref
81632B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 81632B is a single-channel optical power sensor module designed for integration into HP/Agilent lightwave measurement systems. Built around an InGaAs sensor element, it delivers precise optical power measurements across the 800 nm to 1650 nm wavelength range with ±3% accuracy in the 1200 nm to 1630 nm band. The module achieves exceptional linearity of ±0.02 dB ±20 pW across its full dynamic range from +10 dBm to -60 dBm, ensuring reliable characterization of optical components and communication systems. Measurement speed reaches 200 microseconds, enabling rapid power assessment workflows. Low polarization dependent loss and typical polarization-related uncertainty of ±0.015 dB minimize measurement errors across all signal states. Output is provided via BNC analog connector, allowing direct integration with external measurement equipment and data acquisition systems.

Technical Specifications

Sensor Type: InGaAs

Wavelength Range: 800 nm to 1650 nm

Power Range: +10 dBm to -80 dBm

Accuracy: ±3% from 1200 nm to 1630 nm

Power Linearity: ±0.02 dB ±20 pW (+10 dBm to -60 dBm)

Relative Uncertainty due to Polarization: Typ. ±0.015 dB

Polarization Dependent Loss: Low PDL

Measurement Speed: 200 microseconds

Output: BNC Analog Output connector

Key Features

Single-channel configuration for dedicated power measurement

InGaAs sensor optimized for near-infrared wavelengths

High linearity performance across dynamic range ensures minimal distortion

Fast measurement acquisition supports high-throughput testing

Low polarization sensitivity reduces environmental variables

Typical Applications

Optical power characterization in telecom and datacom systems

Component-level testing of optical transceivers and modules

Lightwave system calibration and verification

Fiber-optic signal monitoring

Compatibility & Integration The 81632B integrates with HP/Agilent 8163A, 8164A, 8166A, and 8163B mainframes and multichannel measurement systems. The module is obsolete and no longer manufactured by Keysight.

Specifications

ParameterValue
Power Range (dBm)'+10 dBm to -80 dBm | Wavelength Range: 800 nm to 1650 nm

HP / Agilent 81632B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 81632B. Calibration covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 81632B faults: FC/PC connector damage, EEPROM failure (mainframe shows unrecognized sensor), reduced responsivity. Common faults include FC/PC input connector damage, EEPROM failure causing the module to appear unrecognized by the mainframe, and reduced detector responsivity from contamination.

HP / Agilent 81632B price

This HP / Agilent 81632B is listed at $1,259.8. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 81632B calibration cost

NIST-traceable calibration for the HP / Agilent 81632B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. Turnaround is 1 to 3 business days after receipt. See full calibration details →