Advantest Q82216 – Optical Sensor Head
Advantest Optical Power Sensors. Power Range (dBm): -94 dBm, Channels: 2 (A and B), Accuracy: ±2.5% at calibration point.
- Model
- Advantest Q82216
- Calibration
- Available on request
- Price
- $600
- Status
- In stock
- RFQ ref
- Q82216
This unit
The Advantest Q82216 is a dual-channel optical sensor head for precision optical power measurement and component testing. It delivers ±2.5% accuracy at calibration point with automatic wavelength sensitivity compensation, dual measurement modes (CW and 270 Hz±0.2% chopped light), and onboard data storage for up to 400 measurements per channel. The unit integrates a visual 11-dot level meter, configurable averaging (2–256×), maximum hold functionality, and GPIB data access, making it suitable for demanding laboratory and production test environments.
Technical Specifications
- Channels: 2 (A and B)
- Measurement Modes: CW or Chopped light at 270 Hz±0.2%
- Accuracy: ±2.5% at calibration point; ±4.5% broadband wavelength region with sensitivity curve compensation
- Linearity: ±0.5%
- Sampling Speed: Up to 100 times/sec
- Ranging Speed: 20–500 ms maximum
- Data Storage: 400 points per channel in backup memory
- Data Interface: GPIB
- Averaging: 2–256× running average (configurable)
- Noise Level: -94 dBm (sensor dependent)
- Optical Fiber Compatibility: 10 µm core diameter, NA ≤0.19
- Connector Options: FC, SC, ST, MU, LC, and plug-in types
- Display: 11-dot level meter
- Dimensions: 10 × 8 × 8 in
- Weight: 5 lbs
Key Features
- Dual-channel simultaneous measurement with independent data logging
- Automatic wavelength compensation eliminates manual recalibration across spectrum
- Selectable CW or modulated light measurement modes for diverse optical sources
- Onboard maximum hold captures transient peak power events
- Real-time visual feedback via dot-matrix display
- High sampling rate (100/sec) enables dynamic characterization
Typical Applications
- Optical component qualification (couplers, splitters, filters)
- Fiber-optic system commissioning and troubleshooting
- Wavelength-dependent power profiling
- Production line power verification
- Modulated optical signal measurement
Compatibility & Integration The Q82216 integrates with the Advantest Q8221 optical power measurement system series. Bare fiber and high return loss adapters (A08020, A08328, A08021) accommodate diverse fiber terminations. Calibration options available through OPT82216+25 and equivalent specifications. Operating Environment: 0 to +40°C, ≤85% humidity, up to 2000 m altitude. Storage: -25 to +70°C.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '-94 dBm |
Advantest Q82216 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Advantest Q82216. Calibration covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Advantest Q82216 faults: FC/PC connector damage, EEPROM failure (mainframe shows unrecognized sensor), reduced responsivity. Common faults include FC/PC input connector damage, EEPROM failure causing the module to appear unrecognized by the mainframe, and reduced detector responsivity from contamination.
Advantest Q82216 price
This Advantest Q82216 is listed at $600. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Advantest Q82216 calibration cost
NIST-traceable calibration for the Advantest Q82216 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. Turnaround is 1 to 3 business days after receipt. See full calibration details →