Tektronix DM5110 DMM Programmable Digital Multimeter

The Tektronix DM5110 is a programmable digital multimeter designed for accurate and reliable measurements in various electronic testing and measurement applications. It features a wide range of measurement functions, including DC/AC voltage, DC/AC current, resistance, and diode testing. Its programmable interface allows for automated testing and data acquisition, making it suitable for both benchtop and system integration.

Model
Tektronix DM5110
Calibration
Available on request
Price
$305
Status
In stock
RFQ ref
DM5110
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Tektronix DM5110 is a 4.5-digit programmable digital multimeter designed for precise electronic measurements in automated test systems and laboratory environments. This modular plug-in for the TM5000 system delivers IEEE-488.1 (GPIB) programmability, enabling remote control and seamless integration into larger test configurations. With autoranging capability, multiple measurement modes, and built-in comparison functions, the DM5110 handles demanding measurement tasks from DC voltage to temperature across a wide dynamic range.


Technical Specifications

Measurement Functions & Ranges:

  • DC Voltage: up to 1000 V, 10 mV resolution

  • AC Voltage (True RMS): up to 500 V, 10 mV resolution, 50 kHz frequency response – 20–100 Hz: ±(0.8% reading + 0.05% F.S.) – 100 Hz–10 kHz: ±(0.3% reading + 0.05% F.S.) – 10–20 kHz: ±(0.6% reading + 0.05% F.S.) – 20–50 kHz: ±(1.0% reading + 0.05% F.S.)

  • DC Current: up to 2000 mA, 10 nA resolution

  • AC Current (True RMS): up to 2000 mA, 10 nA resolution

  • Resistance: up to 20 MΩ, 10 mΩ resolution

  • Temperature: −62 °C to +240 °C

  • dBm/dBV calculations via software Display & Indicators:


4.5-digit LED display (±19,999 counts) with NULL, HOLD, AUTO, ADDR, and REM indicators. Basic DC Accuracy: ±0.05% (DC Voltage 2 V range: ±(0.1% reading + 0.02% F.S.))

Key Features



  • Autoranging with manual range override

  • IEEE-488.1 GPIB interface (SH1, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, E1 functions)

  • Null offset measurement capability

  • Hold mode for value retention

  • Triggered measurement operations

  • Compare mode with user-selectable HI/LO limits

  • Beeper alarm for limit exceedance

  • FAST (3.5-digit) and NORMAL (4.5-digit) measurement modes

  • 50/60 Hz line frequency measurement modes

  • Front/rear panel input selection for LOW and Volts/Ω/Temp inputs


Typical Applications Automated testing, data acquisition, production environments, and laboratory measurement systems requiring programmable control and high measurement accuracy.


Compatibility & Integration Modular TM5000 system plug-in with full IEEE-488.1 GPIB programmability for system integration.

Specifications

ParameterValue
Frequency Range50 Hz to 50 kHz (AC measurements) | Voltage / Current Range: DC Voltage: 0–1000 V
AC Voltage0–500 V
DC Current0–2000 mA
AC Current0–2000 mA

Tektronix DM5110 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Tektronix DM5110. Calibration covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Tektronix DM5110 faults: blown input fuse, worn input jacks, display segment failure, relay contact wear on bench DMMs. Common faults include worn input jacks (banana socket contact wear on the mA and COM terminals), blown input fuses (especially the 440 mA HRC fuse), and display segment failures. Keithley 2000/2001/2002 bench DMMs commonly develop relay contact wear affecting resistance accuracy after extended service.

Tektronix DM5110 price

This Tektronix DM5110 is listed at $305. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Tektronix DM5110 calibration cost

NIST-traceable calibration for the Tektronix DM5110 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers DC voltage accuracy at all ranges, AC voltage (True RMS) accuracy at 50 Hz, 1 kHz, and 10 kHz, DC current accuracy, resistance accuracy at all decade ranges, frequency accuracy, and diode test voltage. Fluke 87V and 289 also include capacitance and temperature calibration.. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents