Advantest R8340A Ultra High Resistance Meter
Advantest Insulation Resistance Testers. Current: 10 fA to 19.999 mA.
- Model
- Advantest R8340A
- Calibration
- Available on request
- Price
- $2,579
- Status
- In stock
- RFQ ref
- R8340A
This unit
The Advantest R8340A Ultra-High Resistance Meter is a precision instrument for measuring extremely high resistance values and micro currents across 16 decades. It combines an input amplifier gain-switching I-V converter, high-speed sampling A/D converter, and low-noise voltage generation to enable sensitive detection of ion currents, leakage currents, and insulation material characterization. The meter measures resistance from 10 Ω to 3 × 10¹⁶ Ω and current from 10 fA to 19.999 mA, with sampling rates up to 100 samples per second.
Technical Specifications
Measurement Range- Resistance: 10 Ω to 3 × 10¹⁶ Ω
- Current: 10 fA to 19.999 mA Voltage Source
- DC supply: 0.000 V to 1000.0 V continuously variable
- Floating measurement capability: Up to 1100 VDC
- Voltage generator: 5 mVp-p noise @ 1000 V, 3 ms rise time (0 to 1000 V), 10 W class
- Source/sink capability with selectable current limits Measurement Performance
- Sampling rate: Up to 100 samples per second
- Input amplifier gain switching for adjustable measurement speed and accuracy
- Automatic range level selection: 20000, 2000, or 200 (UP mode); 1799, 179, or 17 (DOWN mode) Data Storage & Output
- Memory capacity: 1000 data items
- Sequence program function with stored measurement conditions
- GPIB interface (IEEE 488-1978) with packed format output support
- BCD output format
- D/A output: ±1 V with ±50% offset adjustment
Key Features
- Programmable measurement sequences including JIS C5102 procedures
- Contact check function via capacitance measurement
- Rapid sample charging and discharging with source/sink capability
- Low-noise, high-response voltage generation
- Configurable input resistance and integration time
Typical Applications
- Insulation material characterization
- Switch and connector testing
- Capacitor evaluation
- Ion current and leakage current detection
- Quality control and production testing
- Research and development
Compatibility & Integration
- Handler interface for external devices and automated fixtures
- GPIB connectivity for test system integration and production line automation
- Incoming inspection and high-volume testing support via packed data format output
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 0.000 V to 1000.0 V DC (source), 10 fA to 19.999 mA, 1100 VDC (floating) |
Advantest R8340A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Advantest R8340A. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Advantest R8340A price
This Advantest R8340A is listed at $2,579. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Advantest R8340A calibration cost
NIST-traceable calibration for the Advantest R8340A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →