Rohde & Schwarz SMY43 9KHz to 2.08GHz Signal Generator
The Rohde & Schwarz SMY43 is a versatile signal generator covering a frequency range of 9 kHz to 2.08 GHz. It provides high-quality signals for a variety of applications, including component testing, receiver testing, and signal simulation. The SMY43 features excellent signal purity and stability, ensuring accurate and reliable results. It offers comprehensive modulation capabilities, including AM, FM, PM, and pulse modulation, making it suitable for a wide range of communication standards and test scenarios.
- Model
- Rohde & Schwarz SMY43
- Calibration
- Available on request
- Price
- $4,480
- Status
- In stock
- RFQ ref
- SMY43
This unit
The Rohde & Schwarz SMY43 is a synthesized RF signal generator spanning 9 kHz to 2.08 GHz, engineered for component testing, receiver validation, and signal simulation across diverse communication standards. The instrument delivers exceptional signal purity with SSB phase noise of -114 dBc/Hz at 1 GHz carrier frequency and 10 kHz offset, coupled with low residual FM and high spurious rejection. Output ranges from -140 dBm to +13 dBm, with optional boost to +25 dBm via the SMY-B40 module.
Technical Specifications
Frequency Performance- Frequency range: 9 kHz to 2.08 GHz
- Frequency resolution: 1 Hz
- Setting time: < 60 ms
- Reference frequency standard (optional SMY-B1): aging of 1 × 10⁻⁶/year after 30 days operation Output Level
- Output range: -140 dBm to +13 dBm
- Overrange capability: +19 dBm standard; +25 dBm with optional SMY-B40
- Output impedance: 50 Ω
- Non-interrupting level setting available over 20 dB range Modulation Capabilities
- Modulation types: AM, FM, PM, Pulse Modulation
- Internal modulation generator: 1 Hz to 500 kHz
- Optional AF synthesizer (SMY-B2): 10 Hz to 100 kHz or 1 Hz to 500 kHz
- FM deviation: up to 20 MHz for VCO simulation; 2 MHz bandwidth for FSK and telemetry applications
- Stereo channel separation: 50 dB
- External AM monitoring and channel switchover logic integrated Signal Quality
- SSB phase noise: -114 dBc/Hz (1 GHz carrier, 10 kHz offset)
- Low residual FM and high spurious rejection Modulation Input
- Input impedance: 100 kΩ; 600 Ω jumper-selectable
Key Features
- Frequency resolution to 1 Hz enables precise tuning across nine decades
- Dual LCD displays for real-time frequency, level, and modulation parameter monitoring
- High-stability optional reference standard with 1 × 10⁻⁶/year aging specification
- AM, FM, and phase modulation with internal AF synthesis to 500 kHz
Typical Applications
- RF component characterization and acceptance testing
- Receiver sensitivity and selectivity measurement
- Signal simulation for system-level validation
- FSK and telemetry signal generation
- VCO performance evaluation
Compatibility & Integration IEEE 488 (IEC 625) IEC-Bus connector enables remote operation and integration into automated test systems. Power supply accepts 100–120 V or 200–240 V AC at system frequencies of 47–440 Hz.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 2080 MHz | Power Range (dBm): '-140 dBm to +19 dBm (up to +25 dBm with option SMY-B40) | Voltage / Current Range: 100–120 V and 200–240 VAC, 47–440 Hz |
Rohde & Schwarz SMY43 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Rohde & Schwarz SMY43. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Rohde & Schwarz SMY43 faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Rohde & Schwarz SMY43 price
This Rohde & Schwarz SMY43 is listed at $4,480. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Rohde & Schwarz SMY43 calibration cost
NIST-traceable calibration for the Rohde & Schwarz SMY43 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →