Agilent/Keysight 59501B D/A CONVERTER
The Agilent/Keysight 59501B is a precision instrument designed for accurate and reliable measurements in electronic testing applications. This unit ensures high performance and dependability for various testing needs.
- Model
- Agilent / Keysight 59501B
- Calibration
- Available on request
- Price
- $110
- Status
- In stock
- RFQ ref
- 59501B
This unit
The Agilent/Keysight 59501B is an isolated digital-to-analog converter and power supply programmer that bridges HP-IB control systems to analog-capable power supplies lacking native bus interface. It enables remote, digitally controlled voltage and current programming of compatible instruments and functions as a standalone low-level DC signal source.
Technical Specifications
Output Voltage Ranges- Unipolar High: 0 to 9.99 V (10 mV resolution)
- Unipolar Low: 0 to 9.999 V (1 mV resolution)
- Bipolar High: -10 to +9.98 V (20 mV resolution)
- Bipolar Low: -1 to +0.998 V (2 mV resolution) DC Output Current: 10 mA maximum Accuracy (at 23°C ±5°C)
- Unipolar High: 0.04% + 0.5 mV
- Unipolar Low: 0.04% + 1 mV
- Bipolar High: 0.04% + 1 mV
- Bipolar Low: 0.04% + 2 mV Temperature Stability
- Unipolar High: 0.01%/°C + 0.5 mV/°C
- Unipolar Low: 0.14%/°C + 0.1 mV/°C
- Bipolar High: 0.013%/°C + 0.5 mV/°C
- Bipolar Low: 0.01%/°C + 0.1 mV/°C Performance Parameters
- D/A Conversion Time: 250 µs
- Output Ripple/Noise: 2 mV rms / 10 mV p-p
- Zero Adjust Range: ±250 mV
- Full Scale Adjust: ±5%
- Isolation: 600 Vdc (HP-IB to output)
- Input Power: 10 VA, 48-63 Hz (100/120/220/240 Vac selectable, -13% to +6%)
Key Features
- Selectable unipolar and bipolar output modes via rear-panel switch
- HP-IB command interface for system integration
- Isolated analog output for safe power supply programming
- Precision zero and full-scale calibration adjustments
- Low output impedance supports 10 mA sourcing capability
Typical Applications Remote voltage and current programming of analog-controlled power supplies in automated test systems. Low-level precision DC signal generation for instrumentation control.
Compatibility & Integration Designed for HP power supplies with analog control inputs but no built-in HP-IB interface. Supports current limit programming when paired with supplies using internal HP-IB options for voltage control.
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | '-1 to +0.998V (Low) |
| Bipolar | -10 to +9.98V (High), Unipolar: 0-9.99V (High) |
Agilent / Keysight 59501B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 59501B. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Agilent / Keysight 59501B price
This Agilent / Keysight 59501B is listed at $110. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight 59501B calibration cost
NIST-traceable calibration for the Agilent / Keysight 59501B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →