HP / Agilent 83525A RF plug-in for 8350A/B, 0.01-8.4 GHz
The HP/Agilent 83525A is an RF plug-in module designed for use with the 8350A/B sweep oscillators. It operates in the frequency range of 0.01 to 8.4 GHz, providing a versatile solution for RF signal generation and testing applications.
- Model
- HP / Agilent 83525A
- Calibration
- Available on request
- Price
- $626
- Status
- In stock
- RFQ ref
- 83525A
This unit
The HP/Agilent 83525A is an RF plug-in module for the 8350A/B sweep oscillator mainframe, delivering signal generation across 0.01 to 8.4 GHz. This dual-band instrument combines wide frequency coverage with leveled output control, internal crystal markers, and full HP-IB programmability to support RF testing, device characterization, and precision frequency measurements.
Technical Specifications
Frequency Range: 0.01 to 8.4 GHz (10 MHz to 8.4 GHz) – Lower Band: 0.01 to 2.1 GHz – Upper Band: 2.0 to 8.4 GHz – Band switch point: 2.0–2.1 GHz with 80 dB with Option 002 (70 dB attenuator) – Power level accuracy: ±1.5 dB – Output flatness: ±1 dB (lower band)
Frequency Accuracy (25°C): ±12 MHz across full band
Sweep Performance: – Minimum sweep time (full band): 17 ms – Power sweep capability for single-test response measurements
Crystal Markers: Internal 10 MHz and 50 MHz markers (below 2 GHz) – 1 MHz marker width: ±100 kHz – 10 MHz marker width: ±200 kHz – 50 MHz marker width: ±300 kHz – LED indicator for marker coincidence
Frequency Reference Output: 1 V/GHz nominal; 10 mV rear panel BNC output
Key Features
Dual-band architecture with automatic switching for continuous 0.01–8.4 GHz coverage
Leveled power output with slope compensation for high-frequency loss mitigation
External marker input for amplitude or Z-axis marking synchronized to sweep frequency
HP-IB programmable with 0.1 dB front panel digital resolution for power setting
Compatible with HP 8755 Swept Frequency Response Test Set for enhanced measurement capability
Typical Applications RF device characterization, frequency response testing, swept-frequency measurements, and precision calibration across microwave bands.
Compatibility & Integration Designed as a direct plug-in for HP/Agilent 8350A/B mainframes. Full remote control via HP-IB interface. Internal crystal frequency references and external marker inputs enable integration with legacy and modern test systems.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 0.01 to 8.4 GHz | Power Range (dBm): '+13 dBm minimum leveled output |
HP / Agilent 83525A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 83525A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP / Agilent 83525A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
HP / Agilent 83525A price
This HP / Agilent 83525A is listed at $626.38. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 83525A calibration cost
NIST-traceable calibration for the HP / Agilent 83525A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →