HP / Agilent 83640B Synthesized Swept-Signal Generator, 0.01 – 40 GHz,+10 dBm calibrated output

HP / Agilent RF Signal Sources. Frequency Range: ** 0.01 GHz to 40 GHz, Resolution: ** 1 kHz standard.

Model
HP / Agilent 83640B
Calibration
Available on request
Price
$12,500
Status
In stock
RFQ ref
83640B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The HP/Agilent 83640B is a synthesized swept-signal generator delivering precision frequency synthesis across 0.01 GHz to 40 GHz with +10 dBm calibrated output power. Built on indirect synthesis architecture, this instrument achieves superior single-sideband phase noise performance while maintaining broad spectral coverage. The 83640B supports multiple operational modes

CW, step sweep, list mode, manual sweep, ramp sweep, and analog sweep variants - alongside modulation capabilities including pulse, AM, and FM. Frequency resolution reaches 1 Hz with Option 008, stepping down to 1 kHz standard. Output power ranges from -20 dBm to +10 dBm leveled; Option 001 extends the range to -110 dBm minimum. Switching time of 10 ms typical (without attenuator change) enables rapid frequency transitions. Temperature stability holds 0.01 dB/°C for output power. This generator is engineered for radar system characterization, wireless communication testing, and microwave circuit validation where signal purity and dynamic range matter.

Technical Specifications

Frequency Range: 0.01 GHz to 40 GHz

Calibrated Output Power: +10 dBm

Output Power Range (Levelled): -20 dBm to +10 dBm; -110 dBm to +10 dBm with Option 001

Single-Sideband Phase Noise: Typically -80 dBc at 10 GHz, 10 kHz offset

Harmonics: Typically -50 dBc below 26.5 GHz; -40 dBc minimum above 26.5 GHz

Frequency Resolution: 1 kHz standard; 1 Hz with Option 008

Switching Time: 10 ms typical (without attenuator change); 15 ms + (step size/1 GHz) × 5 ms for frequency band steps

Temperature Stability (Output Power): 0.01 dB/°C typical

Source Match (Internally Leveled): 1.6:1 SWR <20 GHz; 1.8:1 SWR <40 GHz; 2.0:1 SWR <50 GHz

Key Features

Indirect synthesis for low-noise performance across the band

Modulation support: pulse, AM (DC to 250 kHz, 3 dB), FM

Pulse modulation for frequencies ≥400 MHz; internal pulse generator included

Option 002 adds 30 ns delay and ±5 ns pulse compression for external pulse inputs

External ALC (Automatic Level Control) supported

Power meter mode at 0.7 Hz nominal

Operating temperature range: 0 °C to 55 °C

Typical Applications

Radar system testing and characterization

Wireless communication signal validation

Microwave circuit design and verification

Compatibility & Integration Multiple sweep modes and modulation types enable integration into automated test systems and manual measurement setups. Option 001 step attenuator expands dynamic range to -110 dBm minimum output.

Specifications

ParameterValue
Frequency Range0.01 GHz to 40 GHz | Power Range (dBm): '-110 dBm to +10 dBm

HP / Agilent 83640B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 83640B. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP / Agilent 83640B faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

HP / Agilent 83640B price

This HP / Agilent 83640B is listed at $12,500. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP / Agilent 83640B calibration cost

NIST-traceable calibration for the HP / Agilent 83640B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents