Agilent/Keysight 86240A

Agilent / Keysight RF Signal Sources. Frequency Range: 2 GHz to 8.6 GHz, Sensitivity: ±20 MHz/Volt, Bandwidth: 10 MHz.

Model
Agilent / Keysight 86240A
Calibration
Available on request
Price
$1,500
Status
In stock
RFQ ref
86240A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Agilent/Keysight 86240A is a RF plug-in module for HP 8620 Series sweep oscillator mainframes, delivering 40 mW of internally leveled output across the 2 GHz to 8.6 GHz frequency band. Purpose-built for applications demanding precise RF power control and signal purity, this module combines stable frequency generation with sophisticated modulation capability and low residual noise.


Technical Specifications

Frequency Range: 2 GHz to 8.6 GHz RF Output Power: 40 mW, internally leveled Output Impedance: 50 Ω nominal Source SWR: <1.6 (internally leveled); typically 3 (unleveled) Signal Purity:

  • Linearity: Better than 0.5%

  • Residual FM: <9 kHz peak (20 Hz – 15 kHz bandwidth, FM switch in NORM, CW mode) External FM Modulation:

  • Maximum deviation (DC to 100 Hz): ±100 MHz

  • Maximum deviation (90 kHz to 10 MHz): ±1.5 MHz

  • Nominal sensitivity: ±20 MHz/Volt

  • Frequency response (DC to 10 MHz): ±1.5 dB

  • FM bandwidth: 10 MHz Output Connector: Type N female


Key Features



  • Internal leveling and slope control for stable power delivery

  • Wide external FM capability spanning DC to 10 MHz bandwidth

  • Dual SWR performance modes for flexible impedance matching

  • Compact form factor: 2.3 kg net weight


Typical Applications Precision RF signal generation, frequency-agile testing, modulation characterization, and power-leveled sweep applications in the X-band region.


Compatibility & Integration Designed exclusively as a plug-in module for HP 8620 Series mainframes. Part of the broader HP 86200 series family, which collectively covers 10 MHz to 22 GHz.

Agilent / Keysight 86240A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 86240A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Agilent / Keysight 86240A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Agilent / Keysight 86240A price

This Agilent / Keysight 86240A is listed at $1,500. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / Keysight 86240A calibration cost

NIST-traceable calibration for the Agilent / Keysight 86240A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents