Anritsu MG3696B Synthesized Signal Generator, 2 to 65GHz
Anritsu MG3696B Synthesized Signal Generator, 2 to 65GHz,Offers high-purity signal generation for microwave and millimeter-wave component and system testing.
- Model
- Anritsu MG3696B
- Calibration
- Available on request
- Price
- $34,530
- Status
- In stock
- RFQ ref
- MG3696B
This unit
The Anritsu MG3696B is a synthesized signal generator delivering high-purity signals from 2 to 65 GHz, extending to 67 GHz operationally. It supports optional low-frequency extension down to 10 MHz, or DC coverage to 0.1 Hz with Option 22. Purpose-built for microwave and millimeter-wave component and system testing, the MG3696B serves R&D, manufacturing, and ATE environments with configurable architecture and exceptional spectral purity.
Technical Specifications
Frequency Coverage- Base range: 2 to 65 GHz operational to 67 GHz
- Optional extension: 10 MHz to 65 GHz (Options 4 or 5)
- DC extension: 0.1 Hz to 65 GHz (Option 22 with Options 4 or 5)
- Extended range with external multipliers: Up to 325 GHz or beyond
- Frequency resolution: 0.01 Hz standard; 0.02 Hz below 10 MHz (Option 22) Output Power
- Standard: +17 dBm typical at 20 GHz
- Leveled output: -120 dBm to +17 dBm in 0.01 dB steps
- High power options: +25 dBm to 10 GHz, +23 dBm to 20 GHz, +19 dBm to 40 GHz, +13 dBm to 50 GHz, +3 dBm to 65 GHz
- Typical high power levels: +26 dBm at 20 GHz, +21 dBm at 40 GHz, +16 dBm at 50 GHz
- 2 dB power degradation with Option 22 Phase Noise & Spectral Purity
- Ultra-low SSB phase noise option: Typically –110 dBc/Hz at 1 kHz offset, 10 GHz carrier
- Standard phase noise: –94 dBc/Hz at 1 kHz offset
- Excellent harmonics and spurious response across full spectrum Switching Performance
- Frequency switching: Typically 5 ms for 1 GHz; internal pulse generator with swept delay for moving target simulation
- IF up-conversion option for IQ modulation solutions
- Menu-driven front panel interface; remote control via IEEE-488 GPIB with SCPI support
- Modulation and analog sweep unavailable in 0.1 Hz to 10 MHz band with Option 22
Typical Applications Component characterization, system integration testing, receiver sensitivity and dynamic range assessment, frequency response mapping, modulated signal generation for transceiver validation.
Compatibility & Integration Full GPIB remote control with native interface language and SCPI programming. Local front-panel operation via intuitive menu navigation.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 2 to 65 GHz (operational to 67 GHz) |
| up to 325 GHz or beyond with external multipliers | Power Range (dBm) | '-120 dBm to +17 dBm (standard) |
Anritsu MG3696B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu MG3696B. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu MG3696B faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Anritsu MG3696B price
This Anritsu MG3696B is listed at $34,530. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu MG3696B calibration cost
NIST-traceable calibration for the Anritsu MG3696B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →