HP / Agilent 83550A RF plug-in, 8.0-20.0 GHz, for 8350B
The HP/Agilent 83550A is an RF plug-in designed for use with the 8350B sweep oscillator. It operates in the frequency range of 8.0 to 20.0 GHz, providing a versatile solution for various RF testing and measurement applications.
- Model
- HP / Agilent 83550A
- Calibration
- Available on request
- Price
- $1,762
- Status
- In stock
- RFQ ref
- 83550A
This unit
The HP/Agilent 83550A is an RF plug-in for the 8350B sweep oscillator mainframe that extends microwave testing capabilities across 8.0 GHz to 20.0 GHz. This unit delivers leveled output power exceeding +20 dBm from 8.0 GHz to 18.6 GHz, dropping to +18 dBm minimum in the 18.6–20.0 GHz band. Output level accuracy is ±1.5 dB with frequency accuracy of ±20 MHz.
Technical Specifications
Frequency Range: 8.0 GHz to 20.0 GHz
Output Leveled Power: >+20 dBm (8.0–18.6 GHz); >+18 dBm (18.6–20.0 GHz)
Output Level Accuracy: ±1.5 dB
Frequency Accuracy: ±20 MHz
Harmonic Spurious Signals: –20 dBc at maximum specified power
Non-Harmonic Spurious Signals: –50 dBc at maximum specified power
Output Connector: N(f)
Dimensions: 12 × 10 × 8 in
Weight: 10 lbs
Key Features
Internal AM modulation with external pulse, AM, and FM modulation options
Built-in source module interface enables operation as a microwave driver for HP 8355X series millimeter-wave source modules
HP-IB programmability through 8350B mainframe for automated sweep control, marker frequencies, and output power adjustment
Typical Applications
Component characterization and receiver testing
System integration and RF/microwave subsystem validation
Scalar and network analyzer measurements
Noise figure and power meter applications
Compatibility & Integration The 83550A integrates with the 8350B mainframe and interfaces with industry standard instruments: network analyzers (HP 8510, HP 8410C), scalar network analyzers (HP 8756A, HP 8757A), counters (HP 5343A), noise figure meters (HP 8970), and power meters (HP 432A/B/C, HP 436A, HP 438A). HP-IB programmability enables integration into automated test systems.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 8.0 GHz to 20.0 GHz | Power Range (dBm): >+20 dBm (8.0–18.6 GHz) |
HP / Agilent 83550A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 83550A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP / Agilent 83550A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
HP / Agilent 83550A price
This HP / Agilent 83550A is listed at $1,762. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 83550A calibration cost
NIST-traceable calibration for the HP / Agilent 83550A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →