Anritsu 69059A Synthesized CW Generator, 10 MHz – 26.5 GHz

Anritsu 69059A,"Synthesized CW generator delivering signals from 10 MHz to 26.5 GHz with high stability and low phase noise. Ideal for RF testing in aerospace, communications, and research environments. Provides precise frequency and power control, GPIB interface for automation, and solid construction for both lab and production use."

Model
Anritsu 69059A
Calibration
Available on request
Price
$17,670
Status
In stock
RFQ ref
69059A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Anritsu 69059A Synthesized CW Generator delivers stable, low-phase-noise RF signal generation across 10 MHz to 26.5 GHz. Engineered for aerospace, telecommunications, and R&D environments, this instrument provides precise frequency and power control with GPIB automation capability. Output ranges from -125 dBm to +17 dBm with phase noise of -100 dBc at 10 kHz offset from 10 GHz carrier.


The 69059A supports CW and step sweep modes - stepping through up to 10,000 frequencies with dwell times from 1 ms to 99 seconds and switching speeds under 40 ms. Frequency resolution reaches 1 kHz standard or 0.1 Hz with Option 11. Programmable frequency agility enables 1000 non-sequential phase-locked steps via GPIB control.


Internal AM, FM, phase, and pulse modulation are standard; complex modulation can be user-downloaded. Frequency accuracy tracks the internal or external 10 MHz time base, with aging stability below 2×10⁻⁸/day and temperature stability below 2×10⁻⁸/°C over 0°C to 55°C. Harmonic suppression achieves <-50 dBc in the 500 MHz to 2.2 GHz band, <-40 dBc from 50 MHz to 2 GHz, and <-30 dBc from 10 MHz to 50 MHz.


Technical Specifications



  • Frequency Range: 10 MHz to 26.5 GHz

  • Output Power: -125 dBm to +17 dBm maximum

  • Phase Noise: -100 dBc at 10 kHz offset (10 GHz carrier)

  • Spurious Signals: <-50 dBc (500 MHz–2.2 GHz); <-40 dBc (50 MHz–2 GHz); <-30 dBc (10–50 MHz)

  • Frequency Resolution: 1 kHz (0.1 Hz with Option 11)

  • Frequency Stability (Aging): <2×10⁻⁸/day (<5×10⁻¹⁰/day with Option 16)

  • Frequency Stability (Temperature): <2×10⁻⁸/°C over 0°C–55°C (<2×10⁻¹⁰/°C with Option 16)

  • Step Sweep Range: 1 to 10,000 steps; 1 kHz to full range step size

  • Dwell Time: 1 ms to 99 seconds per step

  • Frequency Switching Time: <15 ms + 1 ms/GHz or <40 ms, whichever is less

  • Programmable Agility: 1000 non-sequential frequencies via GPIB


Key Features



  • CW and step sweep modes with alternate sweep capability

  • Manual sweep with user-selectable step count or size

  • Internal AM, FM, phase, and pulse modulation

  • User-downloadable complex modulation

  • GPIB (IEEE-488) remote control and automation

  • Volatile memory for frequency list storage


Typical Applications



  • Aerospace and defense RF testing

  • Telecommunications subsystem characterization

  • Research laboratory signal generation

  • Production test automation


Compatibility & Integration GPIB interface enables seamless integration into automated test systems and allows real-time frequency and power parameter adjustment under remote control.

Specifications

ParameterValue
Frequency Range10 MHz to 26.5 GHz | Power Range (dBm): '-125 dBm to +17 dBm

Anritsu 69059A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Anritsu 69059A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Anritsu 69059A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

Anritsu 69059A price

This Anritsu 69059A is listed at $17,670. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Anritsu 69059A calibration cost

NIST-traceable calibration for the Anritsu 69059A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →