HP / Agilent 81622B Ge Optical Head, 5mm Diameter
HP / Agilent Optical Power Sensors. Accuracy: ±3.0% (950–1630 nm), Resolution: 0.001 dB.
- Model
- HP / Agilent 81622B
- Calibration
- Available on request
- Price
- $2,653
- Status
- In stock
- RFQ ref
- 81622B
This unit
The HP/Agilent 81622B is a Germanium optical sensor head designed for precision optical power measurement across the infrared spectrum. This 5 mm diameter detector integrates with Agilent optical power meter systems via interface modules such as the 81618A or 81619A, delivering accurate readings from 850 nm to 1650 nm. The sensor achieves ±3.0% accuracy across 950 nm to 1630 nm and offers exceptional linearity of ±0.05 dB ±40 nW in the 950–1630 nm band. Built for demanding laboratory and industrial applications, the 81622B accepts standard single-mode and multi-mode fiber up to 100 µm core diameter (NA = 0.3) and delivers 0.001 dB power resolution with ±0.02 dB repeatability.
Technical Specifications
Sensor Element: Germanium (Ge)
Optical Diameter: 5 mm
Wavelength Range: 850 nm to 1650 nm
Power Measurement Range: +27 dBm to −55 dBm (1250–1650 nm); +23 dBm to −55 dBm (850–1650 nm)
Measurement Accuracy: ±3.0% (950–1630 nm)
Linearity (CW, 23°C ±5°C): ±0.05 dB ±40 nW (+27 dBm to −40 dBm, 950–1630 nm)
Power Resolution: 0.001 dB
Measurement Repeatability: ±0.02 dB
Noise (peak-to-peak): < 40 nW
Polarization Dependence: 45 dB
Maximum Input Power: 300 mW
Optical Damage Threshold: 500 mW/cm²
Analog Output: 0–2 V via BNC connector
Key Features
Germanium detector optimized for near-infrared and infrared wavelengths
Supports standard single-mode and multi-mode fiber (core up to 100 µm, NA = 0.3)
Real-time analog output for instantaneous signal monitoring
High return loss (> 45 dB) minimizes reflective artifacts
Compact 80 × 42 × 190 mm form factor; 0.4 kg
Typical Applications
Optical power measurement in fiber optic systems
Infrared detector calibration and validation
Laboratory characterization of broadband and narrowband sources
Industrial optical power monitoring
Compatibility & Integration The 81622B integrates with Agilent optical power meter platforms via the 81618A or 81619A interface modules. Optical input adapts to common fiber connectors (FC/PC, SC, ST) or bare fiber through available connector adapters. Threaded or D-type magnetic interface options are available depending on model variant.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '+27 dBm to −55 dBm (1250–1650 nm) |
| +23 dBm to −55 dBm (850–1650 nm) | Voltage / Current Range | 0–2 V (analog output) | Wavelength Range: 850 nm to 1650 nm |
HP / Agilent 81622B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 81622B. Calibration covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP / Agilent 81622B faults: FC/PC connector damage, EEPROM failure (mainframe shows unrecognized sensor), reduced responsivity. Common faults include FC/PC input connector damage, EEPROM failure causing the module to appear unrecognized by the mainframe, and reduced detector responsivity from contamination.
HP / Agilent 81622B price
This HP / Agilent 81622B is listed at $2,653. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 81622B calibration cost
NIST-traceable calibration for the HP / Agilent 81622B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity across all specified wavelengths (typically 800-1700 nm range), absolute power accuracy, and EEPROM calibration coefficient update. Compatible with Agilent/HP 8163A/B, 8164A/B, and 8166A/B mainframes.. Turnaround is 1 to 3 business days after receipt. See full calibration details →