LeCroy DA1822A Differential Amplifier, 10MHz

Lecroy Signal Conditioning Modules. Bandwidth: DC to 10 MHz, Accuracy: ±1% + termination resistance uncertainty, Accuracy: ±0.05%.

Model
Lecroy DA1822A
Calibration
Available on request
Price
$3,979
RFQ ref
DA1822A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The LeCroy DA1822A is a 10 MHz differential amplifier that conditions low-level differential signals for measurement on single-ended instruments including oscilloscopes, network analyzers, and spectrum analyzers. It delivers selectable gain from 1 to 1000 with a built-in input attenuator, enabling sensitivity enhancement down to 2 µV/div and precise characterization of power delivery networks and high-current shunt resistor voltages.


Technical Specifications



  • Bandwidth: DC to 10 MHz

  • Gain Settings: 1, 10, 100, or 1000 (selectable)

  • Input Attenuator: Selectable ÷1 or ÷10, yielding effective gains of 0.1, 1, 10, 100, or 1000

  • Risetime: < 35 ns at ×1 or ×10 gain

  • Maximum Output: ±5 V into 50 Ω load

  • Output Impedance: 50 Ω

  • Gain Accuracy: ±1% + termination resistance uncertainty

  • Attenuator Accuracy: ±0.05%

  • Maximum Differential Linear Input: 5 V

  • Equivalent Input Noise Voltage: < 9 nV/√Hz broadband at ×10 gain

  • DC Drift: 50 µV/°C at ×10 gain

  • Common Mode Rejection Ratio: 100,000 to 1 (frequency-dependent)

  • Input Resistance (÷1 Attenuator): 1 MΩ or 100 MΩ

  • Input Resistance (÷10 Attenuator): 1 MΩ

  • Input Capacitance (÷10 Attenuator): 20 pF

  • Input Leakage Current (Attenuator): < 10 pA at 0–45°C

  • Input Protection: ±250 V with automatic disconnect and manual reset

  • Input Range (÷1 Attenuator): ±15.5 V

  • Input Range (÷10 Attenuator): ±155 V


Key Features



  • Built-in Precision Voltage Generator (PVG) with 5½-digit resolution

  • PVG voltage range: ±15.5 V (±10 V in differential offset mode)

  • PVG DC accuracy: 0.05% + 500 µV via temperature-controlled oven

  • Digital control with individual digit increment/decrement

  • Autozero function at 0.0 V selection and periodic intervals

  • PVG output available externally via rear panel connector

  • Configurable as differential comparator input or internal offset voltage source


Typical Applications



  • Measuring voltages across high-current shunt resistors

  • Power structure loss characterization

  • Low-voltage differential signal conditioning for single-ended measurement instruments

  • Extended oscilloscope sensitivity for microvolt-level measurements


Compatibility & Integration Front panel differential inputs accept standard signal sources. Amplifier output and PVG voltage are available via rear panel connectors for instrument integration and external applications.

Specifications

ParameterValue
Frequency RangeDC to 10 MHz | Voltage / Current Range: ±15.5 V (÷1 attenuator), ±155 V (÷10 attenuator)

Lecroy DA1822A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Lecroy DA1822A. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Lecroy DA1822A price

This Lecroy DA1822A is listed at $3,979. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Lecroy DA1822A calibration cost

NIST-traceable calibration for the Lecroy DA1822A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →