Stanford Research FS752 GPS/GNSS Time and Frequency Reference

Stanford Research Systems Clock & Timing Analyzers. Accuracy: <.

Model
Stanford Research Systems FS752
Calibration
Available on request
Price
$3,179
Status
In stock
RFQ ref
FS752
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Stanford Research Systems FS752 is a GPS/GNSS disciplined time and frequency reference engineered for precision synchronization in metrology labs, observatories, and critical infrastructure requiring sub-nanosecond timing accuracy. It tracks GPS, GLONASS, BEIDOU, and GALILEO constellations simultaneously, delivering stable 10 MHz and 1 PPS outputs locked to atomic clock accuracy. The instrument employs a double-oven controlled 3rd overtone SC-cut crystal oscillator with <2×10⁻⁹ temperature stability and phase noise below –125 dBc/Hz at 10 Hz offset.


Satellite acquisition completes within one minute; almanac lock takes approximately 15 minutes under continuous power. If GNSS signal is lost, the holdover timebase maintains lock with aging less than ±0.05 ppm/year.


Technical Specifications

Oscillator:

  • Double-oven controlled, 3rd OT SC-cut crystal

  • Temperature stability: <2×10⁻⁹ (20–30 °C)

  • Aging: <0.2 ppm/year (undisciplined to GPS)

  • Phase noise: <–125 dBc/Hz at 10 Hz offset GNSS Receiver:

  • Constellations: GPS, GLONASS, BEIDOU, GALILEO

  • Satellite acquisition: <1 minute typical

  • Almanac acquisition: ~15 minutes (continuous power)

  • Holdover aging: <±0.05 ppm/year 10 MHz Outputs:

  • Standard: Five buffered outputs (optional: up to 8 additional)

  • Amplitude: 1 Vrms into 50 Ω

  • Harmonics: <–40 dBc

  • Spurious: <–90 dBc (100 kHz bandwidth)

  • Output coupling: DC, 50 Ω ±2 %

  • Reverse protection: ±5 VDC 1 PPS Outputs:

  • Standard: Two buffered outputs (optional: up to 8 additional)

  • Width: 10 µs

  • Phase accuracy: <2 ns (to internal reference)

  • Jitter: <50 ps rms

  • Level: +5 V CMOS logic

  • Transition time: <2 ns

  • Source impedance: 50 Ω

  • Reverse protection: ±5 VDC Stability:

  • TDEV between instruments: A few nanoseconds


Key Features



  • Multi-constellation GNSS tracking eliminates single-system dependency

  • Automatic position survey and fixing

  • Configurable alarm relay for GNSS lock, timebase health, or holdover monitoring

  • USB interface for full programmability and real-time status queries

  • Holdover operation maintains frequency lock when GNSS signal drops


Typical Applications Precision frequency distribution, laboratory timing synchronization, network timing infrastructure, frequency metrology, and applications requiring <2 ns phase accuracy.


Compatibility & Integration USB control port supports command-driven operation. Standard 90–264 VAC input, 47–63 Hz. Power consumption <80 VA. No power switch; instrument activates on line cord insertion.

Specifications

ParameterValue
Frequency Range10 MHz | Voltage / Current Range: 47–63 Hz

Stanford Research Systems FS752 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Stanford Research Systems FS752. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Stanford Research Systems FS752 price

This Stanford Research Systems FS752 is listed at $3,179. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Stanford Research Systems FS752 calibration cost

NIST-traceable calibration for the Stanford Research Systems FS752 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents