Exfo FTB-7600E-0034B SM OTDR Long-Haul Module 1550/1625 nm 50/48dB

Exfo OTDRs. Dynamic Range: 50 dB at 1550 nm, Resolution: Up to 128.

Model
Exfo FTB-7600E-0034B
Calibration
Available on request
Price
$10,930
Status
In stock
RFQ ref
FTB-7600E-0034B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Exfo FTB-7600E-0034B is a single-mode OTDR module built for long-haul fiber optic network characterization. Operating at 1550 nm and 1625 nm with dynamic ranges of 50 dB and 48 dB respectively, it delivers precise fault location, splice loss assessment, and bend detection across extended fiber routes. The module achieves event dead zones below 1 m and attenuation dead zones below 5 m, enabling accurate analysis of closely spaced anomalies. With sampling resolution to 128,000 points and up to 256,000 data points, it supports distance measurements to 256 km and reflectance detection down to −70 dB.

Technical Specifications Optical Performance

Wavelengths: 1550 nm, 1625 nm

Dynamic Range: 50 dB at 1550 nm; 48 dB at 1625 nm

Pulse Widths: 5 ns to 10 µs (variable)

Sampling Resolution: Up to 128,000 points

Data Points: Up to 256,000 points

Distance Range: Up to 256 km

Event Dead Zone: < 1 m (typical)

Attenuation Dead Zone: < 5 m (typical)

Reflectance Measurement: Down to −70 dB (typical)

Key Features

Dual-wavelength OTDR operation for comprehensive link analysis

Bi-directional testing enables averaging for improved splice loss and attenuation accuracy

Configurable loss thresholds and distance ranges

Support for chromatic dispersion (CD) and polarization mode dispersion (PMD) measurements when paired with appropriate FTBx platforms and modules

Integration with digital talk sets and video inspection probes for connector end-face assessment

Typical Applications

Fault location in long-haul fiber routes

Splice loss quantification and assessment

Macrobend and break detection

Connector characterization

Bi-directional fiber link validation

Compatibility & Integration

Compatible with EXFO FTB-2, FTB-2 Pro, and FTB-400 platforms

Fiber Type: Single-mode (SMF)

Optical Connectors: FC/UPC, SC/UPC, LC/UPC (interchangeable)

Data Interface: USB for data transfer and firmware updates

Specifications

ParameterValue
Power Range (dBm)'-70 dBm | Wavelength Range: 1550 nm, 1625 nm

Exfo FTB-7600E-0034B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Exfo FTB-7600E-0034B. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Exfo FTB-7600E-0034B faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.

Exfo FTB-7600E-0034B price

This Exfo FTB-7600E-0034B is listed at $10,930. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Exfo FTB-7600E-0034B calibration cost

NIST-traceable calibration for the Exfo FTB-7600E-0034B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents