Keysight N9323C Basic Spectrum Analyzer, 1 MHz to 13.6GHz

Keysight Spectrum Analyzers. Frequency range: 1 MHz to 13.6 GHz, Accuracy: ±(0.22% × span + span/(sweep point −1)), Bandwidth: 10 Hz to 3 MHz in 1-3-10 sequence.

Model
Agilent N9323C
Calibration
Available on request
Price
$8,500
RFQ ref
N9323C
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Keysight N9323C Basic Spectrum Analyzer is a general-purpose RF measurement instrument spanning 1 MHz to 13.6 GHz. It delivers rapid spectrum analysis with swept and FFT modes, supporting frequency spans from 0 Hz (zero span) to 13.6 GHz. Resolution bandwidth ranges from 10 Hz to 3 MHz in 1-3-10 sequence, with typical RBW accuracy of ±5%.


The analyzer achieves a displayed average noise level of −144 dBm at 1 GHz, enabling sensitive measurements across demanding test scenarios.


Technical Specifications

Frequency Performance

  • Frequency range: 1 MHz to 13.6 GHz

  • Frequency span: 0 Hz to 13.6 GHz

  • Frequency span accuracy: ±(0.22% × span + span/(sweep point −1))

  • Sweep time: 2 ms to 1000 s (span >100 Hz)

  • Resolution bandwidth: 10 Hz to 3 MHz in 1-3-10 sequence

  • RBW accuracy: ±5% nominal

  • Resolution filter shape factor: <5:1 nominal (60 dB/3 dB bandwidth ratio) Amplitude and Dynamic Range

  • DANL @ 1 GHz: −144 dBm typical

  • Typical DANL: −154 dBm/Hz @ 1 GHz

  • Maximum DANL: −162 dBm

  • Measurement range (1 MHz–500 MHz): DANL to +10 dBm (preamp off)

  • Measurement range (500 MHz–13.6 GHz): DANL to +20 dBm (preamp off)

  • Absolute amplitude accuracy: ±0.3 dB at 50 MHz (typical)

  • Overall amplitude accuracy: ±1.3 dB (1 MHz–7 GHz)

  • Input attenuator: 0 to 50 dB in 5 dB steps

  • Input attenuation switching uncertainty: ±0.2 dB typical (at 50 MHz)

  • RBW switching uncertainty: +0.1 dB typical

  • Third order intercept: +11 dBm to +15 dBm across bands

  • Marker level readout resolution: 0.01 dB Phase Noise

  • Phase noise @ 1 GHz (10 kHz offset): −92 dBc/Hz

  • Phase noise @ 500 MHz (10 kHz offset): −92 dBc/Hz

  • Phase noise @ 1 MHz offset: −119 dBc/Hz Input Protection and Triggering

  • Maximum damage level: +30 dBm average continuous (input attenuator ≥20 dB)

  • Maximum damage level: +33 dBm or ±50 VDC

  • DC voltage maximum: ±50 VDC

  • Trigger types: RF Burst, External, Video

  • Level display units: dBm, dBmV, dBμV, W, V, and EMF variants


Key Features



  • Compact spectrum analysis from 1 MHz to 13.6 GHz

  • Configurable resolution bandwidth and sweep parameters

  • Multiple trigger modes for automated measurements

  • Extensive amplitude display formats

  • Input protection to ±50 VDC and +33 dBm continuous power


Typical Applications



  • EMI troubleshooting and compliance testing

  • Spectrum monitoring and signal detection

  • Radio link monitoring

  • Wireless IoT device characterization

  • Educational RF and microwave laboratories


Compatibility & Integration The N9323C accepts RF signals via standard connector with programmable attenuation, external triggering, and video trigger modes for flexible test automation.

Specifications

ParameterValue
Frequency Range1 MHz to 13.6 GHz | Power Range (dBm): −162 dBm to +20 dBm | Voltage / Current Range: ±50 VDC maximum

Agilent N9323C — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent N9323C. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Agilent N9323C price

This Agilent N9323C is listed at $8,500. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent N9323C calibration cost

NIST-traceable calibration for the Agilent N9323C is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →