Fluke 2640A NetDAQ Networked Data Acquisition Unit
Fluke Data Acquisition Systems (DAQ). Channels: 10 (addition, Current: ** 4–20 mA loops, Voltage: 300 VDC/VAC rms (Channels 1 & 11).
- Model
- Fluke 2640A
- Calibration
- Available on request
- Price
- $1,650
- Status
- In stock
- RFQ ref
- 2640A
This unit
The Fluke 2640A NetDAQ is a networked data acquisition unit that combines universal analog inputs, digital I/O, and thermocouple measurement capability in a single distributed system. It acquires data from multiple sensor types simultaneously and transmits results across Ethernet networks, enabling remote monitoring and control across industrial and laboratory environments. Expandable to 400 channels using up to twenty units in networked configuration.
Technical Specifications
Channel Capacity
Analog Inputs: 20, expandable to 400 across networked systems
Computed Channels: 10 (addition, subtraction, multiplication, division, logarithm, exponentiation, square root, absolute value, integer function, average)
Digital I/O & Alarm Outputs: 8 total
Totalizer: 1
Measurement Ranges & Accuracy
DC Volts: 90 mV to 300 V, resolution 0.3 µV to 1 mV, ±0.01% accuracy (3-Sigma) AC Volts: 300 mV to 300 V, resolution 10 µV to 10 mV, ±0.3% accuracy (3-Sigma) Resistance: 300 Ω to 10 MΩ, resolution 300 mΩ to 3 MΩ, ±0.015% accuracy (3-Sigma) Frequency: 15 Hz to 1 MHz, resolution 0.01 Hz to 100 Hz, ±0.05% accuracy (3-Sigma) Temperature (Pt 100 RTD): −200 to 600 °C, resolution 0.003 °C, ±0.06 °C accuracy (3-Sigma) Temperature (J Thermocouple): −100 to 760 °C Current: 4–20 mA loops
Performance
Data acquisition rate: up to 1,000 readings/second system capacity; 2640A model delivers 100 measurements/second with 5½ digit resolution and 0.02% accuracy
Slow Scan: 6 readings/second
Medium Scan: 41–48 readings/second (50/60 Hz)
Fast Scan: 143 readings/second
A/D Converter: multi-slope type, 18-bit linear resolution to 0.3 µV and 0.02 °C
Common Mode Rejection: ≥120 dB (AC and DC)
Normal Mode Rejection: 50 dB @ 50/60 Hz
Maximum Common Mode Voltage: 300 VDC/VAC rms (Channels 1 & 11); 150 VDC/VAC rms (all others)
Key Features
Universal Input Module accepts any combination of input types without external signal conditioning
Ethernet (IEEE 802.3, 10Base2/10BaseT, TCP/IP) for remote access and control
Multi-range capability with automatic range selection
Typical Applications
Multi-point temperature and voltage monitoring in distributed environments
Industrial process data logging and control systems
Laboratory measurement networks requiring high precision and expandability
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 15 Hz to 1 MHz | Voltage / Current Range: 90 mV to 300 V DC |
Fluke 2640A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Fluke 2640A. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Fluke 2640A price
This Fluke 2640A is listed at $1,650. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Fluke 2640A calibration cost
NIST-traceable calibration for the Fluke 2640A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →