Keithley 7011-S Quad 1x10 Mux Switch Card
Keithley Modular Instrument Chassis. Current: 16 mA per relay, Voltage: 175 V peak (any pin to chassis), Current: <.
- Model
- Keithley 7011-S
- Calibration
- Available on request
- Price
- $325
- Status
- In stock
- RFQ ref
- 7011-S
This unit
The Keithley 7011-S is a Quad 1x10 Multiplexer Switch Card engineered for integration with Keithley 7001 and 7002 Switch Mainframes. It delivers flexible signal routing through four independent 1x10 2-pole multiplexer banks, configurable as 2-, 4-, or 8-pole units. Banks combine for expanded topologies: dual 1x20, 1x10 with 1x30, single 1x40, or dual 7011-S cards for a 1x80 multiplexer via backplane interconnection.
Technical Specifications Multiplexer Architecture
Four independent 1x10 2-pole banks (Form A: Hi, Lo contacts)
40 total 2-pole channels per card
Configurable pole widths: 2, 4, or 8 poles per bank
Bank isolation via jumper selection
Actuation time: 3 ms
Relay drive current: 16 mA per relay Electrical Performance
DC signals: 110 V DC, 1 A switched, 30 VA (resistive load)
AC signals: 125 V rms / 175 V peak, 1 A switched, 60 VA (resistive load)
Common mode voltage: 175 V peak (any pin to chassis)
Contact potential: <500 nV per channel contact pair
Offset current: <100 pA
Channel resistance: 10⁹ Ω, 10⁹ Ω, 10⁹ Ω, 10⁹ Ω, 10⁹ Ω, 10⁹ Ω, <600 pF
Crosstalk @ 1 MHz, 50 Ω load: <–40 dB (bank and channel)
Insertion loss (50 Ω source/load): <0.1 dB below 1 MHz, <3 dB below 2 MHz Contact Reliability
Cold switching life: 10⁸ closures
Life at maximum signal levels: 10⁵ closures
Key Features
Screw terminal interface (#16 AWG maximum, 0.092 inch O.D.)
Supports #22 AWG typical (88 conductors per card max)
Expandable via 7001/7002 backplane architecture
Bank jumper configuration for flexible isolation
Compatibility & Integration Designed exclusively for Keithley 7001 and 7002 Switch Mainframes with backplane connectivity for multi-card expansion.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | DC to 2 MHz | Voltage / Current Range: 1 A, 1 A |
| AC | 125 V rms / 175 V peak, DC: 110 V |
Keithley 7011-S — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Keithley 7011-S. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Keithley 7011-S price
This Keithley 7011-S is listed at $325. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Keithley 7011-S calibration cost
NIST-traceable calibration for the Keithley 7011-S is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →