Agilent / Keysight 85036B Standard Mechanical Calibration Kit, DC to 3 GHz, Type-N, 75 ohm
The Agilent/Keysight 85036B is a standard mechanical calibration kit designed for calibrating network analyzers and other RF test equipment. It operates from DC to 3 GHz and uses Type-N connectors with a 75-ohm impedance. This kit includes essential calibration standards such as open, short, and load to ensure accurate measurements.
- Model
- Agilent / Keysight 85036B
- Calibration
- Available on request
- Price
- $2,198
- Status
- In stock
- RFQ ref
- 85036B
This unit
The Agilent/Keysight 85036B is a mechanical calibration kit for vector network analyzers operating from DC to 3 GHz with Type-N connectors and 75-ohm impedance. It delivers the essential open, short, and load calibration standards required for accurate RF measurements across this frequency range.
Technical Specifications
- Frequency Range: DC to 3 GHz
- Impedance: 75 ohms
- Connector Type: Type-N
- Connector Torque Setting: 135 N-cm ±13.5 N-cm (12 in-lb ±1.2 in-lb)
Key Features The kit contains male and female open and short standards, a 75-ohm load termination, three Type-N to Type-N 75-ohm adapters, and one female open center-conductor extender. All opens and shorts are machined to precision tolerances. The short's inner conductor features one-piece construction referenced to the shorting plane for repeatability. The female open uses a separate plastic inner conductor with low dielectric constant to minimize compensation requirements. Adapters employ dual-beaded connector geometry for stable, repeatable connections built to tight tolerances. Pin depths are controlled to specification across all standards:
Type-N Male Connectors: 0.207 to 0.210 inch
Type-N Female Connectors: 0.204 to 0.207 inch
Male Short Pin Depth: 0 to -0.0020 inches
Female Short Pin Depth: 0 to -0.004 inches
Type-N Adapters Pin Depth: 0 to -0.0030 inches The kit withstands repeated use and handling without degradation of connection repeatability. Storage includes a foam-lined case with protective end caps.
Typical Applications Calibration of vector network analyzers and RF test equipment requiring accurate S-parameter measurements across the DC to 3 GHz band.
Compatibility & Integration Designed specifically for Agilent/Keysight network analyzers. Use open-end wrenches when disconnecting to prevent damage. Store with end caps and maintain clean mating plane surfaces. Performance verification requires return to Keysight Technologies for recertification using specialized equipment.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | DC to 3 GHz |
Agilent / Keysight 85036B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 85036B. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Agilent / Keysight 85036B price
This Agilent / Keysight 85036B is listed at $2,197.78. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight 85036B calibration cost
NIST-traceable calibration for the Agilent / Keysight 85036B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →