Agilent / Keysight 8683A Signal Generator

Agilent / Keysight RF Signal Sources. Voltage / Current Range: -10%), <200 VA max, 100/120/220/240 VAC (+5%, 48–66 Hz, Sensitivity: -5 MHz/V.

Model
HP 8683A
Calibration
Available on request
Price
$12,778
RFQ ref
8683A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Agilent/Keysight 8683A is a discontinued RF signal generator engineered for component testing, receiver validation, and calibration procedures. It delivers stable, accurate signal generation with precise frequency and amplitude control in a portable form factor suitable for benchtop and field maintenance use.


Technical Specifications

Modulation

  • AM modulation: 0–70% depth at 1 kHz rate

  • FM modulation: Internal FM with 1 kHz nominal sweep rate and variable deviation to ±5 MHz (D models extend to ±10 MHz in x2 bands)

  • Pulse modulation with internal generator: rate 10 Hz to 1 MHz (5 ranges), width 50 ns to 100 ms (7 ranges), delay 100 ns width, ON >+1.0 V peak, OFF <+0.4 V peak

  • RF pulse rise/fall time: 80 dB; minimum width 1 MHz; peak power ±0.5 dB relative to CW level Signal Quality

  • Phase lock input sensitivity: -5 MHz/V Power

  • Selectable 100, 120, 220, or 240 V (+5%, -10%), 48–66 Hz

  • Maximum consumption: <200 VA Environmental

  • Operating temperature: 0° to 55°C

  • EMI compliance: MIL-STD-461, VDE0871, CISPR Pub. 11

  • Safety: IEC 348 Physical

  • Dimensions: 145 mm H × 457 mm W × 472 mm D

  • Weight: 17.9 kg net


Key Features



  • Continuously adjustable pulse parameters across seven ranges for precise temporal control

  • High-speed RF pulse switching with 80 dB on/off isolation

  • Internal FM with fixed sweep rate; extended deviation options in D-series models

  • Remote control via SCPI commands over LAN with local lockout capability

  • Dual-voltage 48–66 Hz operation from universal AC mains


Typical Applications



  • RF component characterization and validation

  • Receiver sensitivity and dynamic range testing

  • Equipment calibration and system commissioning

  • Field service and portable bench measurement


Compatibility & Integration Remote operation via SCPI commands over LAN. Front-panel lockout function disableable by power cycle, LOCAL command, or Preset key.

Specifications

ParameterValue
Voltage / Current Range'-10%), <200 VA max, 100/120/220/240 VAC (+5%, 48–66 Hz

HP 8683A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the HP 8683A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common HP 8683A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.

HP 8683A price

This HP 8683A is listed at $12,778. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

HP 8683A calibration cost

NIST-traceable calibration for the HP 8683A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →