Delta 9076 Precision Temperature Controller
Delta Legacy / Obsolete Test Equipment. Model: 9076.
- Model
- Delta 9076
- Calibration
- Available on request
- Price · starting from
- $1,100
- Status
- In stock
- RFQ ref
- 9076
This unit
The Delta 9076 is a environmental test chamber designed for demanding applications requiring precise temperature control. It has exceptional reliability and is suitable for use in industrial automation, aerospace, and high-end electronics testing.
Technical Specifications * Model: 9076
- Brand: Delta Design (now part of Cohu Semiconductor Equipment Group)
- Type: Environmental Test Chamber / Temperature Chamber
- Test Volume: 5.78 cubic feet (163.7 liters)
- Internal Dimensions: 20"H x 20"W x 25"D (50.80 cm x 50.80 cm x 63.50 cm)
- Overall Dimensions: 36.25"H x 38.25"W x 37.5"D (92.08 cm x 97.16 cm x 95.3 cm)
- Temperature Range: -73°C to +315°C (-100°F to +600°F)
- Optional Temperature Range: Expandable to -184°C (-300°F) with optional LN2 (Liquid Nitrogen) cooling
- Standard Chamber Coolant: 1050 psi LCO2 (Liquid Carbon Dioxide)
- Heat-up Rate: 8°C / minute
- Cool-down Rate: 12°C / minute (Note: Some sources indicate 20°C/minute for a different configuration or model, so verify for specific unit)
- Air Circulation: Horizontal air flow pattern
- Access Ports: Three 7/8" diameter access ports in the rear
Electrical Specifications * Power Requirements (Varied by voltage configuration): * 208 VAC, 50 or 60 Hz, three-wire, single-phase, 4800W * 208 VAC, 50 or 60 Hz, four-wire, single-phase, 4800W * 240 VAC, 50 or 60 Hz, single-phase, 4800W * 120 VAC, 50 or 60 Hz, single-phase, 1800W (for smaller configurations or older models, confirm for 9076)
Control and Interfaces * Temperature Controller: Model 9010 (standard)
- Programming Capability: * Allows programs to be repeated up to 9,999 times * Each program step includes temperature, time, and rate of change parameters * Soak time programmable up to 100 hours * Program parameters entered via front panel buttons with LCD display * Battery-backed memory for program storage * Optional programmer, Model 9015 or 9020, available for more extensive test routines
- External Interfaces: IEEE-488 and RS232
- Temperature Accuracy: * Setpoint to actual: ±1.0°C * Deviation after stabilization: ±0.1°C * Repeatability: ±0.25°C
- Line Voltage Stabilization: ±0.25°C for 105-125VAC or 220-240VAC line voltage change
- Safety Features: Adjustable over-temperature protection device
- Controller Features: Pushbutton convenience, backlit LCD, settable bus address and baud rates
Physical Specifications * Approximate Net Weight: 232 lbs (105.9 Kg)
- Construction: Combination of aluminum and plastic parts for a solid and rigid structure (Note: This reference is for a handler, but indicates general construction principles)
Compatibility and Accessories * Carts: Model 9901 Cart (open framework) works with use with the Model 9076 chamber. Model 9904 Cart works with chambers including the 9076 and LCO2 bottles.
- Doors: Specific door options are available for the 9076, including options with trays and ports. Why Choose Aumictech for the Delta 9076? At Aumictech, we specialize in supplying environmental test chambers. Whether you need precise temperature cycling for component validation, reliable burn-in testing capabilities, or solid equipment for aerospace and industrial applications, our team ensures the Delta 9076 delivers maximum performance and accuracy for your critical testing workflows.
Delta 9076 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Delta 9076. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Delta 9076 price
This Delta 9076 is listed at $1,100. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Delta 9076 calibration cost
NIST-traceable calibration for the Delta 9076 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →