Exfo IQ-1623 2-Channel High-Speed Power Meter
Exfo Optical Power Meters. Channels: 2 (simultaneous), Dynamic Range: Up to 80 dB, Wavelength: dependent power monitoring.
- Model
- Exfo IQ-1623
- Calibration
- Available on request
- Price
- $2,400
- Status
- In stock
- RFQ ref
- IQ-1623
This unit
The Exfo IQ-1623 is a 2-channel high-speed optical power meter engineered for precise measurements across demanding telecommunications, research, and manufacturing environments. Built on the IQ-1600 Series platform, this instrument delivers simultaneous dual-channel monitoring with rapid stabilization and high sampling capability, making it ideal for WDM component characterization, continuous power level tracking, and production line testing.
Technical Specifications Optical Performance
Detector Type: InGaAs
Wavelength Range: 800 nm to 1700 nm
Power Measurement Range: -85 dBm to +6 dBm
Number of Channels: 2 (simultaneous)
Dynamic Range: Up to 80 dB
Polarization-Dependent Response (low-PDR option): ±0.005 dB Sampling & Response
Continuous Mode: Up to 256 samples per second
High-Speed Acquisition Mode: Up to 4096 samples per second
Stabilization Time (automatic gain): Within 15 ms for ≤95 dB fluctuations
Stabilization Time (manual gain): As little as 1 ms Triggering & Control
Trigger Types: Power level trigger; TTL voltage trigger (0 to 5 V)
Channel Synchronization: Simultaneous synchronization of all optical channels Display & User Interface
Graphical interface with configurable measurement displays
Graph Mode supports up to four simultaneous curves
Real-time display in Continuous mode; post-processed display for high-speed acquisitions
Min/max signal tracking capability Physical & Environmental
Dimensions: 12 cm (H) × 3.8 cm (W) × 26.2 cm (D)
Weight: 0.7 kg
Operating Temperature: 0 °C to 50 °C
Storage Temperature: -35 °C to 70 °C
Relative Humidity: 0 % to 80 % non-condensing
Key Features
Dual independent channel architecture enables parallel optical signal analysis
Fast automatic gain switching with 15 ms stabilization across 95 dB dynamic range
Flexible manual gain mode option for sub-millisecond response requirements
TTL and power-level trigger inputs for test synchronization
Compact form factor (0.7 kg) supports bench and system integration
Typical Applications
WDM component testing and validation
Assembly line optical power verification
Continuous wavelength-dependent power monitoring
Multi-channel signal characterization in R&D environments
Compatibility & Integration Designed for integration with the IQS-500/600 Intelligent Test System. Compatible with other IQ Series instruments for expanded test capability.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '-85 dBm to +6 dBm | Voltage / Current Range: 0 to 5 V (TTL trigger input) | Wavelength Range: 800 nm to 1700 nm |
Exfo IQ-1623 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Exfo IQ-1623. Calibration covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Exfo IQ-1623 faults: detector contamination, damaged FC/PC input port, display failure. Common repairs include detector port cleaning or replacement, display repair, and battery or capacitor replacement in older HP/Agilent/Keysight units.
Exfo IQ-1623 price
This Exfo IQ-1623 is listed at $2,400. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Exfo IQ-1623 calibration cost
NIST-traceable calibration for the Exfo IQ-1623 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. Turnaround is 1 to 3 business days after receipt. See full calibration details →