Exfo IQS-1710 1-Channel Power Meter
Exfo Optical Power Meters. Dynamic range: 80 dB, Accuracy: typical polarization-dependent responsivity of ±0.
- Model
- Exfo IQS-1710
- Calibration
- Available on request
- Price
- $1,100
- Status
- In stock
- RFQ ref
- IQS-1710
This unit
The Exfo IQS-1710 is a single-channel optical power meter module engineered for integration into the Exfo IQS-600 Integrated Qualification System. This modular instrument delivers fast, accurate power measurements across laboratory and field environments by pairing with compatible optical heads from the OHS-1700 series. The platform emphasizes high-speed acquisition, flexible display options, and straightforward operation within a standardized test framework.
Technical Specifications Measurement Performance
Continuous-mode peak acquisition speed: up to 5208 acquisitions per second
Sampling rate: up to 5208 Hz
Stabilization time: 300 µs
Dynamic range: 80 dB
Accuracy: typical polarization-dependent responsivity of ±0.008 dB (0 dBm to −50 dBm); wavelength and connector type affect final uncertainty values Data Acquisition
Continuous mode operation
Min/Max acquisition tracking over user-defined timespans
Virtual channels for real-time power ratio calculations and referencing
Graph mode with simultaneous display of up to four curves Electrical
Power measurement unit: configurable (Absolute or Reference mode)
Electronic dark current nulling supported
Key Features
Individually calibrated optical heads enable interchangeability without accuracy loss
User-configurable measurement displays optimized for real-time analysis
Remote control via GPIB (IEEE-488.1, IEEE-488.2), Ethernet, or RS-232 when integrated with IQS-600
Instrument drivers: LabVIEW, SCPI commands, COM/DCOM library drivers
Physical Characteristics Dimensions & Weight
Module dimensions: 125 mm × 36 mm × 282 mm (4 15/16 in × 1 7/16 in × 11 1/8 in)
Module weight: 0.64 kg (1.4 lb) Environmental
Operating temperature: 0 °C to 40 °C (32 °F to 104 °F); maximum 30 °C for optical power > 35 dBm; maximum 25 °C with FOA-396 attenuator
Storage temperature: −40 °C to 70 °C (−40 °F to 158 °F)
Relative humidity: 0 % to 80 % non-condensing
Compatibility & Integration Host system: Exfo IQS-600 Integrated Qualification System. Compatible optical heads (OHS-1700 series) required for operation; sold separately.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | 0 dBm to −50 dBm | Wavelength Range: 1000 nm to 1650 nm |
Exfo IQS-1710 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Exfo IQS-1710. Calibration covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Exfo IQS-1710 faults: detector contamination, damaged FC/PC input port, display failure. Common repairs include detector port cleaning or replacement, display repair, and battery or capacitor replacement in older HP/Agilent/Keysight units.
Exfo IQS-1710 price
This Exfo IQS-1710 is listed at $1,100. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Exfo IQS-1710 calibration cost
NIST-traceable calibration for the Exfo IQS-1710 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. Turnaround is 1 to 3 business days after receipt. See full calibration details →