Agilent / Keysight M9380A PXIe CW Source, 1 MHz to 3 GHz or 6 GHz
Agilent / Keysight RF Signal Sources. Accuracy: Better than ±0.4 dB absolute, Frequency range: Typical Applications.
- Model
- Agilent / Keysight M9380A
- Calibration
- Available on request
- Price
- $3,382
- Status
- In stock
- RFQ ref
- M9380A
This unit
The Keysight M9380A is a modular PXIe continuous wave (CW) source engineered for high-speed automated test system integration. Composed of three integrated PXIe modules - the M9310A source output, M9301A synthesizer, and M9300A frequency reference - this instrument delivers stable, accurate signal generation across 1 MHz to 3 GHz or 6 GHz. Output power reaches +18 dBm to +19 dBm depending on frequency band and manufacturing revision.
Phase noise performance of < −122 dBc/Hz at 1 GHz (20 kHz offset) supports demanding component testing, local oscillator substitution, and interference injection applications.
Technical Specifications
Frequency Coverage: 1 MHz to 3 GHz (standard) or 1 MHz to 6 GHz (optional, via license key upgrade) Output Power: +18 dBm to +19 dBm across frequency range (±0.01 dB settable resolution). Manufacturing revision-dependent specifications apply: units with higher serial numbers support +10 dBm to +19 dBm (1 MHz to 5 GHz) and +10 dBm to +18 dBm (5 GHz to 6 GHz). Earlier revisions deliver +10 dBm to +19 dBm (1 MHz to 2.5 GHz) and +10 dBm to +18 dBm (2.5 GHz to 6 GHz). Amplitude Accuracy: Better than ±0.4 dB absolute; degrades ±0.02 dB/°C outside controlled temperature range (20–30 °C) Phase Noise: < −122 dBc/Hz at 1 GHz (20 kHz offset); −122 dBc/Hz at 10 kHz offset Spectral Purity: −29 dBc harmonics (at 1 GHz); −70 dBc non-harmonics (at 1 GHz) Frequency Switching Speed: 5 ms Frequency Stability: < ±0.5 ppb/day aging (post 72-hour warm-up); < ±0.1 ppm/year Frequency Accuracy (Internal Timebase): ± 10 ppb (controlled temperature); ± 50 ppb (full temperature range). Warm-up: < ±0.1 ppm (5 minutes) or < ±0.01 ppm (15 minutes) over 20–30 °C External Reference: 1 MHz to 110 MHz sine wave input; ±1 ppm nominal lock rangeKey Features
- Compact modular architecture fits PXIe infrastructure
- Fast frequency switching supports multi-channel test sequences
- Low phase noise enables precision LO applications
- Frequency range upgradeable via software license
- VSWR specified across full frequency range
Typical Applications
- Component and subsystem RF testing
- Local oscillator substitution in receiver chains
- Interference and rejection testing
- Stimulus generation for automated measurement systems
Compatibility & Integration Native PXIe form factor integration with high-speed data interfaces enables tight coupling to test platforms. Supports external frequency reference inputs for synchronized multi-instrument operation.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 1 MHz to 3 GHz or 1 MHz to 6 GHz | Power Range (dBm): '+10 dBm to +19 dBm |
Agilent / Keysight M9380A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight M9380A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Agilent / Keysight M9380A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
Agilent / Keysight M9380A price
This Agilent / Keysight M9380A is listed at $3,382. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight M9380A calibration cost
NIST-traceable calibration for the Agilent / Keysight M9380A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →