ANRITSU MW9070B OPTICAL TIME DOMAIN REFLECTOMETER WI FULL SET W CASE
ANRITSU MW9070B OPTICAL TIME DOMAIN REFLECTOMETER WI FULL SET W CASE,Complete OTDR package with full accessory set and protective case. Delivers reliable fiber diagnostics with onboard storage and solid field usability.
- Model
- Anritsu MW9070B
- Calibration
- Available on request
- Price
- $430
- Status
- In stock
- RFQ ref
- MW9070B
This unit
The Anritsu MW9070B is a portable Optical Time Domain Reflectometer (OTDR) engineered for fiber optic installation, commissioning, and maintenance across diverse network architectures. This field-rated instrument delivers automated fault detection with simultaneous waveform trace and event table visualization on a 7-inch semi-transparent LCD display, enabling rapid diagnostics in both indoor and outdoor environments.
Technical Specifications
Physical & Operational- Dimensions: 290 (W) × 120 (H) × 35 (D) mm
- Weight: 3.2 kg (7.05 lbs)
- Battery runtime: >5 hours continuous operation
- Display: 7-inch LCD, semi-transparent for indoor/outdoor visibility
- Construction: Dustproof, field-rated design Measurement Performance
- Dynamic range: Up to 41.5/39.5 dB (with MW0972C optical unit); 36/34 dB (with MW0970B/0972B)
- Dead zone: 3 m Fresnel / 8 m back-scattered (MW0972C); 5 m event / 25 m attenuation (MW0970B/0972B)
- Distance accuracy: ±2 m ± (10⁻⁴ × distance) ± marker resolution
- Loss linearity: ±0.05 dB/dB or 0.1 dB minimum
- Return loss accuracy: ±4 dB
- Threshold range: 0.01 dB to 9 dB in 0.01 dB increments
- Typical measurement time: 45 seconds (5 km @ 1.31 µm, 0.35 dB/km loss, auto settings)
- Maximum auto-measurement time: 180 seconds Safety Classification
- OTDR Output: 21CFR Class 1, IEC Class 1
- Optional Visual LD Light Source: 21CFR Class 2, IEC Class 2
Key Features
- Automated fault location and connector status verification
- High-speed real-time sweep function (0.4 s with MW0970B/0972B)
- Automatic optimization of distance range, pulse width, and averaging parameters
- Event editing and waveform comparison for loss change detection
- Simultaneous display of trace waveform and event table
- Measures transmission loss, connection loss, and return loss
Typical Applications
- Optical trunk line installation and commissioning
- Subscriber line diagnostics
- Optical CATV cable verification
- Optical LAN infrastructure testing and troubleshooting
Compatibility & Integration Six interchangeable optical units support single-mode and multimode fiber at multiple wavelengths: MW0970B (1.31 µm SMF), MW0972B (1.31/1.55 µm SMF), MW0972C (1.31/1.55 µm SMF enhanced performance), MW0973J (0.85 µm GIF), and MW0975J (0.85/1.3 µm GIF).
Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 0.85 µm, 1.3 µm, 1.31 µm, 1.55 µm |
Anritsu MW9070B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Anritsu MW9070B. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Anritsu MW9070B faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Anritsu MW9070B price
This Anritsu MW9070B is listed at $430. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Anritsu MW9070B calibration cost
NIST-traceable calibration for the Anritsu MW9070B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →