Horiba / PTI OC-4000 Optical Controller
The Horiba / PTI OC-4000 Optical Controller is a device designed to control and manage optical components within spectroscopy systems. It enables precise adjustments and synchronization of light sources, detectors, and other optical elements, ensuring optimal performance in various analytical and research applications.
- Model
- Horiba / PTI OC-4000
- Calibration
- Available on request
- Price
- $1,200
- Status
- In stock
- RFQ ref
- OC-4000
This unit
The Horiba PTI OC-4000 Optical Controller provides precise synchronization and phase control of optical choppers and detectors in spectroscopy systems. This unit delivers selectable disk configurations - 5-sector and 30-sector - enabling operation across wide frequency and exposure time ranges. The OC-4000 combines analog external control, TTL synchronization outputs, and phase adjustment capability to support lock-in detection and time-resolved measurements in fluorescence and absorption spectroscopy.
Technical Specifications Frequency Range
5-Sector Disk: 5 Hz to 667 Hz
30-Sector Disk: 30 Hz to 4000 Hz Exposure Time
5-Sector Disk: 750 µs to 100 ms
30-Sector Disk: 125 µs to 17 ms Phase Jitter
5-Sector: 3.0% at 5 Hz, 1.5% at 10 Hz, 0.36% above 20 Hz
30-Sector: 0.18% at 25 Hz, 0.12% at 50 Hz, 0.06% above 100 Hz Linearity: ±0.5% Accuracy: ±0.5% of least significant digit across entire range Phase Control Range: 0 to 180 degrees External Control Input: 0 to 4 VDC TTL Output: 5 Volt standard
Key Features
LED frequency readout for direct monitoring
Wide-range speed control knob for chopper frequency adjustment
Precise phase adjuster for optical signal alignment relative to chopper blade position
Switchable disk configurations (5-sector and 30-sector) for application-specific frequency requirements
RJ-12 connector for chopper drive control
BNC connectors for synchronization and reference signals
Typical Applications The OC-4000 serves lock-in detection systems, time-correlated single-photon counting experiments, and modulated light source synchronization in fluorescence spectroscopy. Its dual-disk design accommodates both low-frequency chopping and high-speed modulation requirements in research and analytical laboratories.
Compatibility & Integration Standard 5 Volt TTL output interfaces directly with lock-in amplifiers, signal averagers, oscilloscopes, and computer data acquisition systems. External control via 0–4 VDC input allows speed regulation from remote equipment. RJ-12 connectivity integrates with PTI fluorescence systems. BNC signal outputs enable straightforward synchronization with detection and analysis instruments.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 30 Hz to 4000 Hz (30-sector disk), 5 Hz to 667 Hz (5-sector disk) | Voltage / Current Range: 0 to 4 VDC (external control input), 5 Volt TTL output |
Horiba / PTI OC-4000 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Horiba / PTI OC-4000. Calibration covers I/O channel accuracy per specification, analog output and input calibration to rated accuracy, and functional communication interface test (serial, Ethernet, fieldbus). All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Horiba / PTI OC-4000 faults: blown I/O fuses, failed analog modules, power supply capacitor failure. Common faults include blown I/O protection fuses, failed analog I/O modules, power supply capacitor failure, and serial or Ethernet communication port damage.
Horiba / PTI OC-4000 price
This Horiba / PTI OC-4000 is listed at $1,200. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Horiba / PTI OC-4000 calibration cost
NIST-traceable calibration for the Horiba / PTI OC-4000 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers I/O channel accuracy per specification, analog output and input calibration to rated accuracy, and functional communication interface test (serial, Ethernet, fieldbus). Turnaround is 1 to 3 business days after receipt. See full calibration details →