ILX Lightwave OMH-6732B – Power/Wavehead

ILX Lightwave Optical Power Meters. Wavelength: dependent detector response, Accuracy: ±3.5%, Accuracy: ±0.5 nm (NIST traceable).

Model
ILX Lightwave OMH-6732B
Calibration
Available on request
Price
$1,250
Status
In stock
RFQ ref
OMH-6732B
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The ILX Lightwave OMH-6732B is a dual-function optical measurement head that combines power and wavelength measurement into a single probe for laser diode and LED characterization. Silicon detectors integrated with an integrating sphere architecture eliminate common measurement obstacles - detector saturation, beam alignment sensitivity, profile variations, polarization effects, and back reflections - delivering repeatable, accurate results across a wide dynamic range. The head self-calibrates wavelength-dependent detector response, temperature-stabilized detectors ensure environmental independence, and all measurements trace to NIST standards.


Technical Specifications

Optical Power Measurement

  • Range: −40 dBm to +30 dBm (100 nW to 1 W)

  • Maximum power: 1 W

  • Accuracy: ±3.5%

  • Linearity: Typical 60 to 70 dB dynamic range Wavelength Measurement

  • Range: 350 nm to 530 nm

  • Accuracy: ±0.5 nm (NIST traceable)

  • Measurement type: Mean wavelength across all spectral content

  • Minimum input power: −4 dBm for wavelengths <1.2 µm Detector

  • Material: Silicon

  • Temperature control: Active stabilization for measurement repeatability


Key Features



  • Integrated power and wavelength measurement eliminates dual-instrument setups

  • Integrating sphere design reduces alignment criticality and polarization sensitivity

  • Wavelength self-calibration removes manual wavelength input requirements

  • Temperature-controlled detector response ensures consistent performance

  • NIST-traceable calibration performed in ILX Lightwave laboratory

  • Linear response across 60–70 dB dynamic range minimizes saturation artifacts


Typical Applications



  • Laser diode parameter characterization and production testing

  • LED spectral and power analysis

  • Optical source validation in integrated photonic systems


Compatibility & Integration Designed for use with the ILX Lightwave OMM-6810B Optical Multimeter or LPA-9080 series laser diode parameter analyzer. Supports FC/PC, FC/APC, SC, ST, DIN connectors and bare fiber configurations via standard adapter assemblies. System-level integration via GPIB/IEEE-488.2 enabled.

Specifications

ParameterValue
Power Range (dBm)−40 dBm to +30 dBm | Wavelength Range: 350 nm to 530 nm

ILX Lightwave OMH-6732B — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the ILX Lightwave OMH-6732B. Calibration covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common ILX Lightwave OMH-6732B faults: detector contamination, damaged FC/PC input port, display failure. Common repairs include detector port cleaning or replacement, display repair, and battery or capacitor replacement in older HP/Agilent/Keysight units.

ILX Lightwave OMH-6732B price

This ILX Lightwave OMH-6732B is listed at $1,250. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

ILX Lightwave OMH-6732B calibration cost

NIST-traceable calibration for the ILX Lightwave OMH-6732B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. Turnaround is 1 to 3 business days after receipt. See full calibration details →