Advantest R3361CN – Spectrum analyzer
The Advantest R3361CN is a spectrum analyzer used for measuring the magnitude of an input signal versus frequency within the full frequency range of the instrument. It is used to analyze the spectral composition of electrical waveforms and other signals.
- Model
- Advantest R3361CN
- Calibration
- Available on request
- Price
- $7,979
- RFQ ref
- R3361CN
This unit
The Advantest R3361CN is a synthesized spectrum analyzer delivering signal analysis from 9 kHz to 2.6 GHz with 1 Hz center frequency resolution. It measures amplitude across a 115 dB display range (−130 dBm to +25 dBm) with 1 dB typical total level accuracy and ±0.3 dB reference level accuracy. Resolution bandwidth spans 30 Hz to 1 MHz in 1 to 3 switchable steps, with selectivity rated at ≤15:1 (60 dB:3 dB).
The instrument incorporates an internal tracking generator outputting 0 to −50 dBm (or 0 to −59 dBm in 1 dB steps) with ±0.5 dB typical accuracy and ≤0.7 dB flatness across 100 kHz to 1.0 GHz. Sweep times range from 50 ms to 1000 s at ±3% accuracy across FREE RUN, LINE, and manual modes. Frequency stability reaches ±2×10⁻⁸/day with drift ≤300 Hz/min after 1 hour warm-up.
Sideband noise measures ≤−105 dBc/Hz at 20 kHz offset. The compact design includes a CRT monochrome display with dual trace capability (screens A and B), AM-FM demodulation, and occupied bandwidth plus adjacent channel leakage power measurement functions. Multi-marker functionality and GPIB (IEC625/IEEE488) interface support field and laboratory deployment.
Technical Specifications
Frequency- Range: 9 kHz to 2.6 GHz
- Center frequency resolution: 1 Hz
- Start/Stop frequency setting supported Amplitude
- Measuring range: −130 dBm to +25 dBm
- Display range: 115 dB
- Total level accuracy: 1 dB (typical)
- Reference level accuracy: ±0.3 dB (typical)
- Maximum input level: +132 dB (with input attenuator ≥30 dB) Resolution Bandwidth
- Range: 30 Hz to 1 MHz
- Switchable in 1 to 3 steps
- Switching accuracy: <10 dB after automatic calibration Selectivity
- Rating: ≤15:1 (60 dB:3 dB) Sweep
- Time range: 50 ms to 1000 s
- Accuracy: ±3% (approximate)
- Modes: FREE RUN, LINE, manual Tracking Generator
- Output level: 0 to −50 dBm or 0 to −59 dBm (1 dB steps)
- Output accuracy: ±0.5 dB at 30 MHz, −10 dBm, 20–30°C
- Output flatness: ≤0.7 dB (100 kHz to 1.0 GHz); ≤1.5 dB (100 kHz to 2.6 GHz) Frequency Stability
- Drift: ≤300 Hz/min (after 1 hour warm-up)
- Reference oscillator: ±2×10⁻⁸/day, ±2×10⁻⁷/year
- Temperature coefficient: ±5×10⁻⁸ (0–50°C, referenced to +25°C)
- Residual FM: ≤20 Hz (Vpp) for 0.1 s (span <2 MHz) Noise
- Sideband noise: ≤−105 dBc/Hz (20 kHz offset, f ≤3.0 GHz) Demodulation
- Type: AM-FM Display
- Type: CRT monochrome
- Dual-trace capability (screens A and B)
Key Features
- Occupied bandwidth measurement
- Adjacent channel leakage power measurement
- Multi-marker function
- Internal tracking generator for frequency characteristic measurement
Interfaces & Integration
- GPIB standard (IEC625 bus, IEEE488)
- RS-232 available (Option 80)
- IC memory card for external storage
Typical Applications RF and microwave signal characterization, component testing, frequency response analysis, and field measurements benefiting from compact, portable design.
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 2.6 GHz | Power Range (dBm): −130 dBm to +25 dBm |
Advantest R3361CN — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Advantest R3361CN. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Advantest R3361CN price
This Advantest R3361CN is listed at $7,979. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Advantest R3361CN calibration cost
NIST-traceable calibration for the Advantest R3361CN is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →