CI Systems SR800R-4D Extended Area Blackbody ControlMaster
The CI Systems SR800R-4D ControlMaster is an extended area blackbody controller designed for precise temperature control and calibration of infrared detectors and systems. It provides highly accurate and uniform thermal radiation for testing and calibration purposes.
- Model
- CI Systems SR800R-4D
- Calibration
- Available on request
- Price
- $5,200
- Status
- In stock
- RFQ ref
- SR800R-4D
This unit
The CI Systems SR800R-4D ControlMaster is an extended area blackbody controller engineered for precise temperature regulation and infrared detector calibration. It delivers highly accurate, uniform thermal radiation across 2µm to 14µm wavelengths with NIST-traceable performance. The system operates standalone or integrates into larger test benches, featuring onboard temperature measurement and digital signal transfer to the controller for enhanced stability and measurement confidence.
Technical Specifications
Temperature Range: -40°C to 1200°C
Temperature Accuracy: 0.007°C; total system uncertainty 0.02°C (T ≤ ±25°C) and 0.03°C (T > ±25°C)
Temperature Stability: Millidegree-Kelvin stability
Typical Yearly Drift: 0.02°C
Set Point and Readout Resolution: 0.001°C
Spectral Emissivity: Greater than 97% (2µm to 14µm)
Emitter Diameter Options: 2" to 20" standard; up to 40" available
Key Features
User-replaceable sensors allow rapid recalibration and system refresh to factory calibration within one year
Optional CK-800R calibration kit enables self-calibration in the field
Millidegree-Kelvin resolution for demanding metrology applications
Large color LCD touchscreen interface with Windows CE operating system
Control of up to four motorized devices from single controller
Built-in MRTD testing and maintenance diagnostics reduce downtime
Compact, portable form factor with included 19" rack-mount kit
Ethernet, GPIB, and RS-232 communication ports for system integration
Remote control software included
Certified to MIL-T-28800D, CE, and FCC standards
Typical Applications
Infrared detector and thermal imaging system calibration
NIST-traceable temperature reference for test facilities
Thermal performance validation and detector characterization
Extended-area radiometric measurement standards
Compatibility & Integration Designed as a standalone unit or for integration into optical test benches with collimators and larger automated test systems. Removable radiation head design supports flexible deployment across multiple test configurations.
Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 2µm to 14µm |
CI Systems SR800R-4D — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the CI Systems SR800R-4D. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
CI Systems SR800R-4D price
This CI Systems SR800R-4D is listed at $5,200. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
CI Systems SR800R-4D calibration cost
NIST-traceable calibration for the CI Systems SR800R-4D is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →