This unit

The Agilent/Keysight 4349A is a precision high resistance meter delivering simultaneous multi-channel measurements for bench-top and manufacturing environments. Available in single-channel, dual-channel, and four-channel configurations, it measures DC resistance from 1 kΩ to 1.6 × 10¹⁶ Ω and DC current from 60 fA to 100 µA with ±0.6% and ±0.4% basic accuracy respectively. Test voltages range from 0.1000 V dc to 1000.0 V dc, enabling characterization of high-impedance devices, surface and volume resistivity, and low-current phenomena.


The instrument delivers measurements in 9 ms typical, with selectable integration times of 28 ms, 98 ms, and 396 ms. Auto and manual ranging, averaging up to 256 samples, and built-in contact detection (2 ms for capacitive devices) ensure reliable production testing. Open error correction removes parasitic effects, while deviation and percent deviation functions support statistical analysis.


Comparator thresholds (high/in/low) and save/recall of up to 10 measurement states streamline test configuration.


Technical Specifications



  • Resistance range: 1 × 10³ Ω to 1.6 × 10¹⁶ Ω

  • Current range: 60 fA to 100 µA

  • Test voltage: 0.1000 V dc to 1000.0 V dc

  • DC current accuracy: ±0.4% basic

  • DC resistance accuracy: ±0.6% basic

  • Measurement time: 9 ms (typical); 28, 98, 396 ms selectable

  • Averaging: 1 to 256

  • Delay time: 0 to 9999 ms (1 ms steps)

  • Test cable length: 2 m maximum

  • Power input: 90–132 V ac or 198–264 V ac, 47–66 Hz; 45 VA typical


Key Features



  • Four-channel simultaneous measurements (Option 700)

  • Contact check capability (2 ms for capacitive devices)

  • Open error correction for parasitic elimination

  • GPIB control (IEEE 488)

  • Handler interface with isolated, negative-logic signals

  • Continuous memory support

  • Compact dimensions: 320 mm (W) × 100 mm (H) × 450 mm (D); 6.5 kg


Typical Applications



  • High-volume resistivity testing (surface and volume)

  • Low-current and high-impedance device characterization

  • Contact resistance verification in manufacturing

  • Electrostatic discharge (ESD) material qualification


Compatibility & Integration Available with 1, 2, or 4 test channels. Four-channel operation requires external DC source. Supports low-noise test leads (16117C, 16117E), resistivity test fixtures (50 mm, 26 mm, 76 mm electrodes), tweezer fixtures, and wide jaw clips.

Agilent / Keysight 4349A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight 4349A. Calibration covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Agilent / Keysight 4349A faults: GPIB interface board failure, display failure, guard terminal switch contact wear, oscillator calibration drift on older analog bridges. HP 4284A and 4285A most common faults are HPIB interface board failure and display contrast loss. The internal oscillator and vector measurement circuits are highly stable and rarely fail. Wayne Kerr and GenRad bridge instruments may develop switch contact wear on decade resistance/capacitance dials.

Agilent / Keysight 4349A price

This Agilent / Keysight 4349A is listed at $1,250. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / Keysight 4349A calibration cost

NIST-traceable calibration for the Agilent / Keysight 4349A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers impedance magnitude accuracy at 100 Hz, 1 kHz, 10 kHz, and 100 kHz, dissipation factor (D) and quality factor (Q) accuracy, phase angle accuracy, open and short compensation verification, and cable length compensation. HP/Keysight 4284A additionally verified at 1 MHz.. Turnaround is 1 to 3 business days after receipt. See full calibration details →