Ametek OCS 500N6.1

Ametek OCS EMI / EMC Test Receivers. Voltage: 250 V to 2, Current: Max. 500 A @ 12 Ω, Voltage: Up to 6 kV.

Model
Ametek OCS 500N6.1
Calibration
Available on request
Price
$11,000
Status
In stock
RFQ ref
500N6.1
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Ametek OCS 500N6.1 is an EMC test instrument that generates damped oscillatory waves and ring waves to evaluate the electromagnetic compatibility of electronic equipment. It simulates transient disturbances on power, control, and signal lines, assessing device robustness under electrical stress conditions. The instrument supports testing per IEC 61000-4-18, IEC 61000-4-12, ANSI/IEEE C62.41, and IEC 61000-4-10 standards, making it essential for laboratory, industrial, and research environments requiring rigorous EMC validation.


Technical Specifications

Damped Oscillatory Waves (DOW)

  • Frequencies: 100 kHz and 1 MHz

  • Open Circuit Voltage (HV output): 250 V to 3,000 V ±10%

  • Line Output Voltage: 250 V to 2,500 V ±10%

  • Rise Time: 75 ns ±20%

  • Source Impedance: 200 Ω ±20% (also 12 Ω or 30 Ω ±20%)

  • Peak Short Circuit Current: Max. 500 A @ 12 Ω; Max. 200 A @ 30 Ω

  • Repetition Rate: Max. 50/s @ 100 kHz; Max. 500/s @ 1 MHz

  • Burst Duration: At least 2 s

  • Polarity: Positive, Negative

  • Decay Ratios: Peak 2 to Peak 1 (0.4–1.1); Peak 3 to Peak 2 (0.4–0.8); Peak 4 to Peak 3 (0.4–0.8); Peak 5 > 50% of Peak 1; Peak 10 < 50% of Peak 1 Ring Wave (RWG)

  • Open Circuit Voltage: Up to 6 kV

  • Rise Time: 0.5 µs ±30%

  • Oscillation Frequency: 100 kHz ±10%

  • Peak Current (Short Circuit): 200–500 A Coupling and Decoupling Network (CDN)

  • Built-in CDN; 1-phase configuration

  • Current Rating: 16 A per line

  • Supply: 250 V / 16 A AC/DC

  • Coupling: Line to line, Line(s) to ground (PE)


Key Features



  • Trigger modes: Automatic, Manual, External

  • Synchronization: 0° to 360° (1° resolution)

  • CRO Trigger output: 5 V signal

  • Control interfaces: USB, GPIB (IEEE 488)

  • Magnetic field testing: External coil compatible (IEC 61000-4-10)


General Specifications



  • Dimensions: 19" / 6 HU

  • Weight: Approx. 28 kg

  • Operating Temperature: 10° C to 35° C

  • Relative Humidity: Max. 85%, non-condensing

  • Atmospheric Pressure: 86 kPa to 106 kPa


Typical Applications EMC compliance testing for power electronics, telecommunications equipment, industrial controls, and signal integrity validation in transient-prone environments.

Specifications

ParameterValue
Frequency Range1 MHz, 100 kHz | Voltage / Current Range: 000 V (DOW HV output)

Ametek OCS 500N6.1 — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Ametek OCS 500N6.1. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Ametek OCS 500N6.1 price

This Ametek OCS 500N6.1 is listed at $11,000. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Ametek OCS 500N6.1 calibration cost

NIST-traceable calibration for the Ametek OCS 500N6.1 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents