HP / Agilent 66321D DC Source 0-15V 0-3A
HP / Agilent DC Power Supplies. Voltage: 0–15 V, Current: 0–3 A, Voltage: ±0.05% of setting + 10 mV.
- Model
- HP / Agilent 66321D
- Calibration
- Available on request
- Price
- $790
- Status
- In stock
- RFQ ref
- 66321D
This unit
The HP/Agilent 66321D is a DC source purpose-built for testing battery-powered wireless communication devices. It delivers 0–15 V and 0–3 A continuous output with 5 A peak current capability (up to 7 ms), making it ideal for characterizing current drain across device operating modes. The unit combines a precision power source with an integrated digital voltmeter and dynamic measurement system, enabling accurate battery behavior simulation and transient load response analysis in production test environments.
Technical Specifications Output Ratings
Voltage: 0–15 V
Continuous current: 0–3 A
Peak current (up to 7 ms): 5 A
Power: 45 W Programming Accuracy (25 °C ±5 °C)
Voltage: ±0.05% of setting + 10 mV
Current: ±0.05% of setting + 1.33 mA DC Measurement Accuracy (25 °C ±5 °C)
Voltage readback: ±0.03% + 5 mV
Current readback: ±0.2% + 0.5 mA (−3 A to +5 A); ±0.1% + 0.2 mA (−1 A to +1 A); ±0.1% + 2.5 µA (−20 mA to +20 mA) Transient Response and Regulation
Transient response time: < 20 µs
Transient voltage drop: < 40 mV
Load regulation: 5 mV voltage, 0.75 mA current
Line regulation: 0.5 mV voltage, 0.75 mA current Ripple and Noise (20 Hz–20 MHz)
Voltage: 1 mV rms, 6 mV peak-to-peak
Current: 2 mA rms
Key Features
Output resistance programming from −0.04 Ω to 1 Ω simulates internal battery resistance (Ni-Cad, Ni-MH, Li-Ion, Li-Polymer)
Negative resistance programming compensates for wiring voltage drops in test fixtures
Dynamic measurement system captures current drain across talk, active, standby, and sleep modes
4096-point buffer with 15 µs sampling interval
GPIB interface for automated test integration
Typical Applications
Current drain characterization of mobile phones, PDAs, and portable wireless devices
Battery life validation for Bluetooth and Wireless LAN products
Transient load testing of cellular handsets and communication modules
Production-volume battery emulation and power delivery analysis
Compatibility & Integration Supports testing of 3G, cdma2000, WCDMA, CDMA, TDMA, GSM, PCS, DECT, TETRA, PHS, NADC, Bluetooth, and Wireless LAN devices. Compatible with Keysight 14565B software for battery drain analysis.
Specifications
| Parameter | Value |
|---|---|
| Voltage / Current Range | 0-15V, 0-3A (peak 5A up to 7ms) |
HP / Agilent 66321D — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 66321D. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
HP / Agilent 66321D price
This HP / Agilent 66321D is listed at $790. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 66321D calibration cost
NIST-traceable calibration for the HP / Agilent 66321D is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →