Agilent/Keysight 71450B – Optical Spectrum Analyzer
Agilent / Keysight Optical Spectrum Analyzers. Power Range (dBm): -90 dBm to +30 dBm, Wavelength Range: 600 to 1700 nm.
- Model
- Agilent / Keysight 71450B
- Calibration
- Available on request
- Price
- $13,529
- Status
- In stock
- RFQ ref
- 71450B
This unit
The Agilent/Keysight 71450B is a diffraction-grating-based optical spectrum analyzer for precise spectral characterization across 600 to 1700 nm. It measures optical power versus wavelength with sensitivity to -90 dBm and delivers 60 dB dynamic range, enabling detailed analysis of optical components, amplifiers, LEDs, DFB lasers, and Fabry-Perot lasers.
Technical Specifications
Wavelength Range: 600 to 1700 nm Sensitivity & Dynamic Range- Sensitivity: -90 dBm
- Dynamic range: 60 dB
- Maximum input power (monochromator): +30 dBm
- Maximum photodetector input level: +20 dBm
- Largest measurable power before compression: +10 dBm Wavelength Accuracy
- Absolute accuracy: ±1 nm
- After user calibration: ±0.3 nm
- After calibration within ±0.05 nm of cal signal: ±0.2 nm
- Span linearity (<40 nm spans): ±0.05 nm
- Span linearity (1530–1570 nm): ±0.02 nm
- Reproducibility (1 minute): ±0.005 nm
- Reproducibility after calibration (1 minute): ±0.001 nm
- Tuning repeatability: ±0.005 nm
- Settability: ±0.005 nm Resolution Bandwidth
- FWHM: <0.08 nm and 0.1 to 10 nm in 1,2,5 sequence
- <0.08 nm setting: 0.065 nm ±15%
- Resolution accuracy (≥0.5 nm, 1250–1600 nm): ±20%
- Resolution accuracy (0.1–10 nm, 600–1700 nm): ±30%
- Corrected bandwidth accuracy for noise markers (≥0.5 nm, 1250–1600 nm): ±3%
- Corrected bandwidth accuracy (0.2 nm, 1250–1600 nm): ±5% Environmental
- Operating temperature: 0° to 55° C
- Relative humidity: <95%
- Stabilization: 1 hour continuous operation plus auto-align routine Physical
- Dimensions: 24 × 20 × 16 in
- Weight: Approximately 35–50 lbs
Key Features
- Unique double-pass monochromator architecture
- Automatic spectral width measurement at -3 dB, -20 dB, or user-defined levels
- Peak power determination and side-mode suppression ratio analysis
- DFB stop-band characterization
- Auto-align routine for sustained amplitude accuracy
- Three instrument modes: preselector, stimulus response, and photodetector test
Typical Applications
- Optical component characterization (couplers, switches, filters, fibers)
- Laser source spectral analysis and quality assessment
- LED spectral profiling
- Stimulus-response measurements for optical devices
Compatibility & Integration Integrates the HP 70950B module for core measurement functionality. Accommodates HP 71451B photodetector for optimized performance across the operating wavelength range.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '-95 dBm to 10 dBm | Wavelength Range: 600 nm to 1700 nm |