HP / Agilent 81689B tunable laser is specially for the C- Band
HP / Agilent Tunable Laser Sources. Resolution: ** 0.01 nm (1.25 GHz at 1550 nm), Accuracy: ** ±0.3 nm, Voltage: ** ±6 V (per BNC connector).
- Model
- HP / Agilent 81689B
- Calibration
- Available on request
- Price
- $7,106
- Status
- In stock
- RFQ ref
- 81689B
This unit
The Agilent 81689B is a compact C-band tunable laser module engineered for optical testing and measurement applications in DWDM systems. It delivers a stable, precise optical source across the entire C-band, enabling flexible test configurations for optical amplifiers, DWDM components, and related optical subsystems. The module supports both internal and external modulation with continuous sweep capability across its full wavelength range, and operates within Agilent's modular lightwave measurement platforms.
Technical Specifications
Wavelength Range: 1525 nm to 1575 nm
Wavelength Resolution: 0.01 nm (1.25 GHz at 1550 nm)
Absolute Wavelength Accuracy: ±0.3 nm
Wavelength Repeatability: ±0.05 nm
Wavelength Stability: ±0.02 nm over 24 hours
Output Power: ≥10 dBm
Output Power Stability: ±0.03 dB over 1 hour; typical ±0.015 dB to ±0.0075 dB over 1 hour at constant temperature
Power Stability (typical): <±0.025 dB over 1 minute
Spectral Linewidth: 20 MHz (Option 021)
Tuning Speed: <10 seconds per 50 nm
Maximum External Voltage: ±6 V (per BNC connector)
Key Features
Tunable across full C-band (1525–1575 nm) with 0.01 nm resolution
Dual modulation capability: internal and external
Continuous sweep through entire wavelength range
Single-mode fiber and polarization-maintaining fiber output options
Straight or angled contact connectors available
Class IIIb laser (FDA 21 CFR 1040.10); Class 3A (IEC 825-1:1993)
Integrated safety circuitry disables optical output on fault detection
NIST-traceable calibration with 2-year recommended recalibration interval
Storage temperature range: -40 °C to +70 °C
Typical Applications
DWDM system characterization and component testing
Optical amplifier performance validation
Wavelength-dependent measurements across C-band
Dynamic wavelength tuning in optical test configurations
Compatibility & Integration The 81689B integrates into Agilent 8163A/B Lightwave Multimeters, 8164A/B Lightwave Measurement Systems, and 8166A/B Lightwave Multichannel Systems. Pairs with the optional 81651A optical attenuator module to extend output power dynamic range.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | ≥10 dBm output | Voltage / Current Range: ±6 V (maximum external) | Wavelength Range: 1525 nm to 1575 nm |
HP / Agilent 81689B — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP / Agilent 81689B. Calibration covers wavelength accuracy at 5 or more points across the full tuning range, mode-hop-free range, output power flatness, linewidth, and side-mode suppression ratio (SMSR). All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP / Agilent 81689B faults: output power drift, mode hops, fiber pigtail connector damage, wavelength inaccuracy. Common faults include output power drift after extended use, mode hops at specific wavelengths, and fiber pigtail connector damage. Full alignment and piezo calibration typically restore performance to spec.
HP / Agilent 81689B price
This HP / Agilent 81689B is listed at $7,106.38. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP / Agilent 81689B calibration cost
NIST-traceable calibration for the HP / Agilent 81689B is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers wavelength accuracy at 5 or more points across the full tuning range, mode-hop-free range, output power flatness, linewidth, and side-mode suppression ratio (SMSR). Turnaround is 1 to 3 business days after receipt. See full calibration details →