HP 8350B and 83592A 10 MHz -20 GHz sweep generator Calibrated
The HP 8350B sweep generator mainframe, combined with the HP 83592A RF plug-in module, creates a high-performance synthesized sweeper covering 10 MHz to 20 GHz. This calibrated unit is ideal for testing a wide range of RF and microwave components and systems.
- Model
- HP 8350B/83592A
- Calibration
- Available on request
- Price
- $2,335
- Status
- In stock
- RFQ ref
- 8350B/83592A
This unit
Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 MHz to 20 GHz | Voltage / Current Range: 100, 120, 220, 50–60 Hz |
HP 8350B/83592A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the HP 8350B/83592A. Calibration covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common HP 8350B/83592A faults: YIG oscillator failure (no output or loss of lock), step attenuator wear (output drops or steps at set attenuation), ALC loop failure (uncontrolled output power), I/Q modulator assembly failure. HP/Agilent 836xx series most common faults are YIG oscillator failure and step attenuator contact wear causing output steps or drops at specific attenuation settings. Keysight E4428C/E4438C units most commonly develop ALC loop failure and I/Q modulator assembly faults. YIG oscillator replacement and attenuator service are stocked repairs.
HP 8350B/83592A price
This HP 8350B/83592A is listed at $2,335. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
HP 8350B/83592A calibration cost
NIST-traceable calibration for the HP 8350B/83592A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers output frequency accuracy, output power level accuracy across the full frequency range, AM/FM/PM/pulse modulation accuracy, harmonic and spurious output levels, and phase noise floor verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →