Agilent/Keysight AC6912L
The Agilent/Keysight AC6912L is a DC power supply, known for its reliability and performance in various testing and measurement applications. This power supply provides stable and precise DC power, essential for powering electronic devices and circuits during development, testing, and maintenance.
- Model
- Agilent / Keysight AC6912L
- Calibration
- Available on request
- Price
- $9,500
- Status
- In stock
- RFQ ref
- AC6912L
This unit
The Keysight AC6912L is a 12 kVA three-phase AC power source that merges power generation with integrated multimeter, harmonic analyzer, and power analyzer functionality. Designed for rigorous electrical system testing, it outputs up to 320 Vrms AC (452 V DC maximum) across configurable single-phase, single-phase three-wire, or three-phase four-wire topologies. The instrument operates from 1 Hz to 550 Hz, supporting AC, DC, and AC+DC modes simultaneously.
It generates custom waveforms, simulates power line disturbances, and measures voltage, current, and power with ±(0.15% + 0.3 V) accuracy across the 45 Hz to 500 Hz band. Built-in harmonic analysis, output sequencing, and programmable soft start/stop and output inhibit functions enable compliance testing for avionics (RTCA DO-160, MIL-STD-704) and general component validation including relays and transformers.
Technical Specifications
- Output Power: 12 kVA
- Output Voltage (Max AC RMS): 320 Vrms
- Output Voltage (Max DC): 452 V
- Output Voltage Setting Range: 0.0–161.0 V (low), 0.0–322.0 V (high)
- Voltage Resolution: 0.1 V
- Voltage Accuracy (45 Hz–500 Hz): ±(0.15% + 0.3 V)
- Frequency Range: 1 Hz to 550 Hz
- Frequency Variation (AC Mode): ±1% (40 Hz to 5 kHz, referenced at 55 Hz)
- Max RMS Current: 120 A / 60 A (1-phase); 40 A / 20 A (1-phase 3-wire or 3-phase)
- Power Factor: 0.95
- Input Voltage: 180–235 V and 380–480 V configurations available
- Input Frequency: 50/60 Hz
- Harmonic Content: 5 Hz to 1 MHz components in DC mode
Key Features
- Waveform generation with custom AC profiles
- Power line disturbance simulation
- Integrated harmonic analysis (Fourier decomposition to 1 MHz)
- Output sequencing for multi-step test protocols
- Programmable soft start/stop and output inhibit
- Over/under-voltage protection
- 5-inch LCD color display
Typical Applications
- Avionics equipment validation (DO-160, MIL-STD-704)
- Power interruption simulation and recovery testing
- Harmonic and voltage fluctuation analysis
- Relay and transformer functional verification
- Electronic system stress and compliance evaluation
Compatibility & Integration
- LAN: LXI Core
- USB: TMC and CDC (Virtual COM)
- GPIB: Optional
Agilent / Keysight AC6912L — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight AC6912L. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Agilent / Keysight AC6912L price
This Agilent / Keysight AC6912L is listed at $9,500. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight AC6912L calibration cost
NIST-traceable calibration for the Agilent / Keysight AC6912L is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →