Yokogawa AQ1210A 1310/1550nm OTDR
The Yokogawa AQ1210A is a compact and lightweight Optical Time Domain Reflectometer (OTDR) designed for testing and troubleshooting optical fiber networks. It operates at 1310nm and 1550nm wavelengths, providing accurate measurements of fiber length, attenuation, and fault location. Ideal for field technicians and network engineers, the AQ1210A ensures reliable fiber optic network maintenance and optimization.
- Model
- Yokogawa AQ1210A
- Calibration
- Available on request
- Price
- $6,279
- Status
- In stock
- RFQ ref
- AQ1210A
This unit
The Yokogawa AQ1210A is a dual-wavelength Optical Time Domain Reflectometer (OTDR) engineered for field characterization and troubleshooting of optical fiber networks. Operating at 1310 nm and 1550 nm with dynamic ranges of 37 dB and 35 dB respectively, it delivers high sampling resolution (5 cm) and short event dead zone (0.75 m) for precise near-end fault detection. The instrument weighs 1 kg and fits the footprint of a tablet, making it portable without sacrificing measurement capability.
Technical Specifications
- Wavelengths: 1310 nm / 1550 nm
- Dynamic Range: 37 dB (1310 nm), 35 dB (1550 nm)
- Event Dead Zone: 0.75 m
- Attenuation Dead Zone: 4 m
- Sampling Resolution: 5 cm
- Maximum Data Points: 256,000
- PON Characterization: Supports up to 1×128 splitters
- Battery Life: 10+ hours
- Battery Recharge Time: 5 hours
- Startup Time: Under 10 seconds
- Power Source: High-capacity Li-Ion polymer battery
Key Features
- 5.7-inch color TFT LCD touchscreen (640×480) with glove-operable capacitive multi-touch and hard-key buttons
- Real-time, averaged, and full auto measurement modes
- Multi-trace, 2-way trace, and differential trace analysis
- Section analysis with return loss measurement
- Smart Mapper function with icon-based network visualization and PASS/FAIL judgment
- Data export: SOR, PDF, JPG, CSV, BMP, SET formats
- Wi-Fi and Ethernet connectivity for remote control and wireless data transfer
- USB Type-A, USB Type-C ports; external fiber inspection probe support
Typical Applications
- PON installation and acceptance testing with high-port-count splitter characterization
- Multi-fiber network verification and fault location
- Event detection in congested fiber sections
- Field reporting with immediate PDF generation and email transmission
Compatibility & Integration Complies with EN61326-1 Class A, EN55011 Class A Group 1, and EN61326-1 Table 2 for industrial environments. Interfaces via Wi-Fi, Ethernet, and USB for integration with field service platforms and cloud storage systems.
Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 1310 nm / 1550 nm |
Yokogawa AQ1210A — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Yokogawa AQ1210A. Calibration covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Yokogawa AQ1210A faults: optical port connector damage or contamination, battery failure, display damage from field use, module failure on FTB or MTS platforms, calibration drift. Most field OTDRs fail due to physical damage. On modular platforms (EXFO FTB, Viavi MTS), individual module failures are serviced without replacing the entire mainframe. Optical port cleaning resolves many apparent measurement faults before mechanical repair is needed.
Yokogawa AQ1210A price
This Yokogawa AQ1210A is listed at $6,279. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Yokogawa AQ1210A calibration cost
NIST-traceable calibration for the Yokogawa AQ1210A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers distance accuracy verified against a calibrated fiber length reference, event loss measurement accuracy at splice and connector events, dead zone (event and attenuation) at rated pulse width, dynamic range verification at 1310 nm and 1550 nm, backscatter coefficient calibration, and return loss measurement accuracy. Turnaround is 1 to 3 business days after receipt. See full calibration details →