Ando AQ6317 600-1750nm Optical Spectrum Analyzer
Ando AQ6317 600-1750nm Optical Spectrum Analyzer,"Broadband optical analyzer offering high-resolution spectrum measurements from 600 nm to 1750 nm. Features fast sweep speeds, low noise floor, and graphical display capabilities. Ideal for testing passive and active optical components in telecom, R&D, and manufacturing environments with precise spectral accuracy."
- Model
- Ando AQ6317
- Calibration
- Available on request
- Price
- $11,570
- Status
- In stock
- RFQ ref
- AQ6317
This unit
Aumictech – Ando AQ6317 Optical Spectrum Analyzer 600–1750 nm Specifications
Optical Performance
Wavelength range: 600 to 1750 nm
Span: 0.2 nm to full range
Wavelength accuracy: ±0.02 nm (typical in C-band), ±0.1 nm across full range
Wavelength resolution: 0.02 nm minimum
Wavelength repeatability: ±0.005 nm
Sampling resolution: 0.001 nm minimum
Level / Power Measurement
Sensitivity (typical):
−85 dBm (1000–1600 nm)
−70 dBm (600–1000 nm)
Maximum safe input power: +20 dBm
Power accuracy: ±0.3 dB (−20 dBm input at 1550 nm)
Linearity: ±0.05 dB (−50 to +10 dBm)
Flatness: ±0.2 dB across 600–1750 nm
Polarization dependence: ≤ ±0.05 dB
Dynamic Range & Stray Light
Dynamic range (typical):
Up to 75 dB (0.05 nm RBW, ±1 nm)
Up to 55 dB (0.02 nm RBW, ±0.2 nm)
Stray light suppression ratio: 75 dB typical
Level range: +20 to −85 dBm
Sweep & Speed
Sweep times (typical, 100 nm span):
FAST mode: ~0.5 s
NORMAL mode: ~2 s
HIGH SENS mode: 30–60 s
Sweep modes: single, repeat, and continuous
Fiber & Optical Interfaces
Input fiber types: Single-mode and multimode (9/125 µm, 50/62.5/125 µm)
Optical input connectors: FC, SC, ST (interchangeable adapters)
Optical return loss: >30 dB typical
Calibration source: built-in wavelength reference
Interfaces & Control
Remote control: GP-IB (IEEE-488), RS-232C standard
Optional interface: Ethernet (for PC control and automation)
Command compatibility: SCPI compliant
Display & Storage
Display: 9-inch color CRT (vector style)
Trace functions: multiple trace storage, subtraction, Max/Min hold
Data output: GP-IB/RS-232C to external PC
Printer support: external GPIB printer compatible
Physical & Power
Dimensions: approx. 425 × 200 × 450 mm (W×H×D)
Weight: approx. 18–20 kg
Power supply: 100–240 VAC, 50/60 Hz, ~150 VA
Operating environment:
Performance guaranteed: +18 to +28 °C
Operating: +5 to +35 °C
Storage: −10 to +50 °C
Humidity: ≤80 %RH non-condensing
Warm-up time: ~1 hour
Why Choose Aumictech for the Ando AQ6317?
The Ando AQ6317 Optical Spectrum Analyzer is a versatile instrument covering 600–1750 nm, making it a go-to tool for telecommunications, photonics R&D, and optical device testing. Its balance of sensitivity, accuracy, and dynamic range makes it one of the most popular legacy OSAs in the field.
At Aumictech, we offer sales, repair, and NIST-traceable calibration services for the Ando AQ6317. Our team ensures this analyzer continues to meet performance requirements, extending the life of legacy test equipment while supporting modern optical research and production testing.
Frequently Asked Questions
- What wavelength range and measurement capabilities does the Ando AQ6317 provide?
- The AQ6317 delivers high-resolution optical spectrum analysis across 600 nm to 1750 nm, covering visible through near-infrared bands essential for telecom and component testing. Its fast sweep speeds and low noise floor enable precise spectral characterization of both passive components like filters and couplers, and active devices including lasers and amplifiers. The instrument's graphical display output makes it straightforward to assess spectral shape, peak wavelength, and signal integrity in R&D and manufacturing environments.
- What is the condition of this used AQ6317, and what has been verified?
- This unit is offered in used condition and has undergone inspection prior to listing. We verify core functionality, display operation, and basic measurement repeatability to ensure the instrument performs as intended for optical spectrum analysis tasks. Calibration status and a formal calibration certificate are available as a separately offered service, starting at $450, allowing you to establish traceability to NIST standards if required for your application.
- What does calibration cover for the Ando AQ6317, and what is the typical cost and turnaround?
- Our calibration service for the AQ6317 covers wavelength accuracy, amplitude response, and noise floor verification across the 600–1750 nm measurement range, with results documented on a certificate of calibration. Service begins at $450, with turnaround dependent on current workload; we recommend contacting us directly for scheduling and detailed pricing. This service restores metrological traceability and confidence in spectral measurements for regulated or high-precision applications.
- What interfaces and software connectivity does the AQ6317 support?
- The AQ6317 features GPIB (IEEE-488) interface for remote control and data transfer, enabling integration into automated test systems and lab networks. Compatible control software and drivers for the AQ6317 are available through Ando documentation; we recommend verifying driver compatibility with your operating system before purchase. USB or Ethernet connectivity options may be available through external interfaces or adapters depending on your specific system configuration.
Ando AQ6317 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Ando AQ6317. Calibration covers wavelength accuracy at 850, 1310, 1550, and 1625 nm, power level accuracy, resolution bandwidth (RBW) accuracy, and dynamic range verification. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common Ando AQ6317 faults: wavelength motor failure, internal light source dimming, detector sensitivity loss. Common repairs include wavelength motor servicing, internal broadband source replacement, and detector recalibration after extended use.
Ando AQ6317 price
This Ando AQ6317 is listed at $11,570. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Ando AQ6317 calibration cost
NIST-traceable calibration for the Ando AQ6317 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers wavelength accuracy at 850, 1310, 1550, and 1625 nm, power level accuracy, resolution bandwidth (RBW) accuracy, and dynamic range verification. Turnaround is 1 to 3 business days after receipt. See full calibration details →