Ando AQ6317 600-1750nm Optical Spectrum Analyzer
Ando AQ6317 600-1750nm Optical Spectrum Analyzer,"Broadband optical analyzer offering high-resolution spectrum measurements from 600 nm to 1750 nm. Features fast sweep speeds, low noise floor, and graphical display capabilities. Ideal for testing passive and active optical components in telecom, R&D, and manufacturing environments with precise spectral accuracy."
- Model
- Ando AQ6317
- Calibration
- Available on request
- Price
- $11,570
- Status
- In stock
- RFQ ref
- AQ6317
This unit
Aumictech – Ando AQ6317 Optical Spectrum Analyzer 600–1750 nm Specifications
Optical Performance
Wavelength range: 600 to 1750 nm
Span: 0.2 nm to full range
Wavelength accuracy: ±0.02 nm (typical in C-band), ±0.1 nm across full range
Wavelength resolution: 0.02 nm minimum
Wavelength repeatability: ±0.005 nm
Sampling resolution: 0.001 nm minimum
Level / Power Measurement
Sensitivity (typical):
−85 dBm (1000–1600 nm)
−70 dBm (600–1000 nm)
Maximum safe input power: +20 dBm
Power accuracy: ±0.3 dB (−20 dBm input at 1550 nm)
Linearity: ±0.05 dB (−50 to +10 dBm)
Flatness: ±0.2 dB across 600–1750 nm
Polarization dependence: ≤ ±0.05 dB
Dynamic Range & Stray Light
Dynamic range (typical):
Up to 75 dB (0.05 nm RBW, ±1 nm)
Up to 55 dB (0.02 nm RBW, ±0.2 nm)
Stray light suppression ratio: 75 dB typical
Level range: +20 to −85 dBm
Sweep & Speed
Sweep times (typical, 100 nm span):
FAST mode: ~0.5 s
NORMAL mode: ~2 s
HIGH SENS mode: 30–60 s
Sweep modes: single, repeat, and continuous
Fiber & Optical Interfaces
Input fiber types: Single-mode and multimode (9/125 µm, 50/62.5/125 µm)
Optical input connectors: FC, SC, ST (interchangeable adapters)
Optical return loss: >30 dB typical
Calibration source: built-in wavelength reference
Interfaces & Control
Remote control: GP-IB (IEEE-488), RS-232C standard
Optional interface: Ethernet (for PC control and automation)
Command compatibility: SCPI compliant
Display & Storage
Display: 9-inch color CRT (vector style)
Trace functions: multiple trace storage, subtraction, Max/Min hold
Data output: GP-IB/RS-232C to external PC
Printer support: external GPIB printer compatible
Physical & Power
Dimensions: approx. 425 × 200 × 450 mm (W×H×D)
Weight: approx. 18–20 kg
Power supply: 100–240 VAC, 50/60 Hz, ~150 VA
Operating environment:
Performance guaranteed: +18 to +28 °C
Operating: +5 to +35 °C
Storage: −10 to +50 °C
Humidity: ≤80 %RH non-condensing
Warm-up time: ~1 hour
Why Choose Aumictech for the Ando AQ6317?
The Ando AQ6317 Optical Spectrum Analyzer is a versatile instrument covering 600–1750 nm, making it a go-to tool for telecommunications, photonics R&D, and optical device testing. Its balance of sensitivity, accuracy, and dynamic range makes it one of the most popular legacy OSAs in the field.
At Aumictech, we offer sales, repair, and NIST-traceable calibration services for the Ando AQ6317. Our team ensures this analyzer continues to meet performance requirements, extending the life of legacy test equipment while supporting modern optical research and production testing.