Agilent / Keysight E1413C C-Size P1 Active Adapter, 1-Slot

The Agilent/Keysight E1413C is a C-size, 1-slot, VXI active adapter designed for GPIB interfacing. It allows seamless integration of GPIB instruments into a VXI system, providing a bridge between the two standards. This adapter ensures reliable communication and control, essential for automated test and measurement applications.

Model
Agilent / Keysight E1413C
Calibration
Available on request
Price
$2,410
Status
In stock
RFQ ref
E1413C
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Agilent/Keysight E1413C is a C-size, 1-slot register-based VXI module that functions as a 64-channel scanning analog-to-digital converter for data acquisition and automated test applications. This module delivers 16-bit resolution across all channels with a maximum aggregate scanning rate of 100 kSamples/second, distributed across the active channel count. A 64 kSample dual-ported FIFO buffer accommodates burst data capture, while on-board calibration sources enable single-command, all-channel calibration with system wiring offset compensation.


Technical Specifications



  • Form Factor: C-Size, 1-Slot, register-based VXI module

  • Channel Count: 64 channels

  • Resolution: 16 bits (including sign)

  • Maximum Scanning Rate: 100 kSamples/second total (divided by number of channels in scan list)

  • Memory Buffer: 64 kSample dual-ported FIFO

  • Accuracy: 0.01% of reading (dependent on SCP module)

  • Input Voltage Ranges: 4 mV to 16 V full-scale (base module); up to 60 V full-scale with E1513A attenuator SCP

  • Maximum Input Voltage: ±16 V peak (normal mode plus common mode)

  • Maximum Common Mode Voltage: ±16 V peak (operating)

  • Input Impedance: >100 MΩ differential

  • Phase Jitter (scan-to-scan): 80 ps RMS

  • Phase Jitter (card-to-card): 41 ns peak, 12 ns RMS

  • Scan Trigger Jitter: ±100 ps typical

  • Internal Scan Trigger Range: 100 µsec to 6.5536 sec, 100 µsec resolution; trigger count 1 to 65535 or infinite

  • External Scan Trigger: TTL level from VXI modules or external hardware

  • Trigger Latency: 17.5 µs typical

  • Sample Timer Range: 10 µsec to 32768 msec, 0.5 µsec resolution

  • Data Format: IEEE 754, 32-bit floating-point

  • VXI Bus Compatibility: A16/A24 Data Transfer Bus, TTL Trigger Bus


Key Features



  • Current Value Table (CVT) stores latest reading from each of 64 channels

  • On-board calibration with system offset compensation

  • TTL trigger output supports multi-module synchronization at same or integer-related rates

  • Supports up to eight mixed Signal Conditioning Plug-on (SCP) modules


Typical Applications Multi-channel data acquisition, automated test systems, sensor scanning, voltage monitoring in production test environments.


Compatibility & Integration The E1413C integrates with Agilent/Keysight SCP modules to customize measurement conditioning. VXI bus connectivity enables integration into larger automated test systems with TTL triggering and synchronization across multiple modules.

Specifications

ParameterValue
Voltage / Current Range4 mV to 16 V typical

Agilent / Keysight E1413C — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight E1413C. Calibration covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Agilent / Keysight E1413C faults: relay contact wear, DAC/ADC drift, VXI backplane connector oxidation. Common faults include relay contact wear, DAC or ADC gain drift, and VXI backplane edge connector oxidation. Most modules are repaired at the component level with full post-repair verification in a C-size or D-size mainframe.

Agilent / Keysight E1413C price

This Agilent / Keysight E1413C is listed at $2,410. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Agilent / Keysight E1413C calibration cost

NIST-traceable calibration for the Agilent / Keysight E1413C is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers all module-specific parameters per manufacturer datasheet, output accuracy and linearity, frequency accuracy, and VXI register and trigger bus compliance. Turnaround is 1 to 3 business days after receipt. See full calibration details →

Documents