The Agilent/Keysight E8364C is a 50 GHz PNA microwave network analyzer engineered for precise S-parameter and RF component characterization across research, development, and manufacturing environments. The instrument integrates a dedicated S-parameter test set with four receivers, enabling true TRL calibration alongside TRM/LRM, on-wafer, in-fixture, waveguide, and antenna measurement modes. Its 8.4-inch LCD touchscreen display and dual-interface operation - hard keys and soft keys - streamline navigation in complex test setups.
Technical Specifications
Frequency and Resolution- Frequency range: 10 MHz to 50 GHz
- Frequency resolution: 1 Hz
- Phase noise: –50 dBc at 1 kHz offset (nominal) Measurement Performance
- Dynamic range: 104 dB (standard, 10 Hz IF bandwidth)
- Trace noise: <0.006 dB RMS at 1 kHz IF bandwidth (standard)
- Test port noise floor: <–114 dBm at 10 Hz IF bandwidth (standard)
- Measurement speed: <26 microseconds per data point
- IF bandwidth range: 1 Hz to 40 kHz (nominal) Power Specifications
- Nominal output power: –12 dBm (standard); –17 dBm with Options 14 and UNL
- Maximum output power: 10 dBm Measurement Channels and Resolution
- Independent channels: 32
- Points per sweep: 2 to 20,001 (user-definable) Calibration Methods
True TRL with four-receiver integration. Supports TRM/LRM, electronic calibration (ECal), user-characterized ECal, adapter removal, unknown Thru, data-based calibration standards, and expanded calibration algorithms.
Key Features
- 2.4 mm male (50 ohm) RF connectors
- Dual GPIB ports (controller and talker/listener)
- USB: 1 front-panel port, 4 rear-panel ports
- 10/100 BaseT Ethernet with auto-switching
- VGA video output (15-pin mini D-Sub)
- Display: 1024×768 resolution touchscreen
- Windows graphical user interface
Typical Applications S-parameter and frequency response analysis. Mixer characterization (conversion loss, return loss, isolation, absolute group delay). Amplifier measurements including gain compression, harmonic and IMD analysis, pulsed-RF testing. Time-domain analysis and antenna characterization.
Compatibility & Integration Operating temperature: 23 °C ± 3 °C for full specification compliance; 25 °C ± 5 °C nominal operation. Relative humidity: 0 to 95% at 40 °C (non-condensing). Altitude to 4600 m (15,000 ft). Optional configurable test set (Option 14) available.
Agilent / Keysight E8364C — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the Agilent / Keysight E8364C. All calibrations are performed to NIST-traceable standards and ship with a certificate. Turnaround is typically 1–3 business days after receipt.
Agilent / Keysight E8364C price
This Agilent / Keysight E8364C is listed at $143,827.9. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
Agilent / Keysight E8364C calibration cost
NIST-traceable calibration for the Agilent / Keysight E8364C is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Turnaround is 1 to 3 business days after receipt. See full calibration details →