ILX Lightwave FPM-8210 Optical Meter
ILX Lightwave Optical Power Meters. Accuracy: ±2.5% (reference conditions), Bandwidth: Key Features.
- Model
- ILX Lightwave FPM-8210
- Calibration
- Available on request
- Price
- $2,080
- Status
- In stock
- RFQ ref
- FPM-8210
This unit
The ILX Lightwave FPM-8210 is a fiber optic power meter designed for precision optical power measurements across the 800 nm to 1650 nm wavelength range. Built around an InGaAs sensor, the instrument delivers ±2.5% accuracy under reference conditions and ±5.0% under operating conditions, with measurement repeatability of ±0.003 dB typical. The single input port accepts both connectorized (FC/PC, FC/APC, LC, SC) and bare fiber configurations, eliminating the need for multiple test heads in production environments.
Technical Specifications
- Wavelength Range: 800 nm to 1650 nm
- Power Measurement Range: -70 dBm to +20 dBm; +23 dBm to –70 dBm for λ < 1000 nm
- Maximum Power Level: +33 dBm at pump wavelengths ( +40 dBm
- Accuracy: ±2.5% (reference conditions), ±5.0% (operating conditions)
- Measurement Repeatability: ±0.003 dB typical
- Polarization Dependent Response: ±0.001 dB typical; ±0.002 dB typical
- Noise: ≤100 pW p-p at 1630 nm typical; <40 pW p-p at 1310 and 1550 nm typical
- Temperature Coefficient: ±0.2%/°C typical
- Linearity: ±0.02 dB, ±100 pW (-60 dBm to +20 dBm)
- Entrance Aperture: 2.54 mm
- Analog Output: 0 to 10 V, 1000 Ω impedance, 0 to 10 Hz bandwidth
Key Features
- InGaAs sensor for extended NIR sensitivity
- Dual measurement range capability for sub-1000 nm and full-band operation
- ±0.003 dB repeatability enables tight process control
- Low polarization sensitivity minimizes calibration complexity
- GPIB (IEEE 488.2) interface for workstation integration
Typical Applications
- Laser and light source characterization
- Optical component test and qualification
- Fiber optic system commissioning
- High-throughput manufacturing test environments
Compatibility & Integration
- Compatible with ILX Lightwave or Agilent bare fiber holders
- CA-120 Bare Fiber Adapter supported
- Standard connector adapters available for FC, SC, and LC interfaces
- Class I equipment; ±10% mains supply voltage tolerance
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | '-70 dBm to +20 dBm | Voltage / Current Range: 0 to 10 V | Wavelength Range: 800 nm to 1650 nm |
ILX Lightwave FPM-8210 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the ILX Lightwave FPM-8210. Calibration covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common ILX Lightwave FPM-8210 faults: detector contamination, damaged FC/PC input port, display failure. Common repairs include detector port cleaning or replacement, display repair, and battery or capacitor replacement in older HP/Agilent/Keysight units.
ILX Lightwave FPM-8210 price
This ILX Lightwave FPM-8210 is listed at $2,080. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
ILX Lightwave FPM-8210 calibration cost
NIST-traceable calibration for the ILX Lightwave FPM-8210 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers detector responsivity at 850, 1310, 1490, 1550, and 1625 nm, absolute power accuracy, linearity across the full dynamic range, and reference power level. Turnaround is 1 to 3 business days after receipt. See full calibration details →