JDS Uniphase JDSU HA9 / HA1 Programmable Attenuator
The JDS Uniphase/JDSU HA9/HA1 is a programmable attenuator designed for precise control of optical signal levels. It features a wide attenuation range, low insertion loss, and high accuracy, making it suitable for various applications, including optical communication systems, test and measurement setups, and research and development.
- Model
- JDS Uniphase/JDSU HA9/HA1
- Calibration
- Available on request
- Price
- $2,930
- Status
- In stock
- RFQ ref
- HA9/HA1
This unit
The JDSU HA9/HA1 Programmable Attenuator series delivers precise optical signal level control for communications, test and measurement, and R&D applications. These units combine wide attenuation range, high accuracy, and low insertion loss to enable critical optical performance evaluation and system calibration. Programmable via GPIB or RS232 with SCPI/HP8156A-compatible commands, the HA9/HA1 supports both local front-panel control and remote operation.
Field-proven construction ensures repeatable performance across demanding test environments.
Technical Specifications
- Attenuation Range: Up to 100 dB (60 dB for HA9W, 750–1700 nm)
- Resolution: HA9 series 0.01 dB; HA1 series 0.001 dB
- Accuracy: ±0.1 dB or greater of ±0.004 dB/dB (1260–1360 nm and 1450–1570 nm); otherwise ±0.1 dB or ±0.015 dB/dB
- Repeatability: 0.01 dB
- Wavelength Range: 1200–1700 nm standard; optional 750–1700 nm
- Wavelength Dependence: 60 dB
- Input Power: Up to 1 W (30 dBm)
- Beam Blocking: Integrated switch provides >90 dB attenuation
Key Features
- Single-mode (9/125 µm) and multimode fiber options
- FC/PC, SC/PC, ST/PC, FC/APC, SC/APC connector adapters
- Front or rear panel mounted universal connector adapters
- 5 V driver output for external 1×1 or 1×2 fiberoptic switch control
- Low spectral ripple design suitable for CATV AM systems
Interfaces & Control
- GPIB (IEEE 488.2) and RS232 remote control
- SCPI/HP8156A command compatibility; software compatible with HP 8157A
- Front panel keyboard for local operation
- Rear panel 5 V output with front panel toggle for external/internal switch selection
Typical Applications Optical communications system calibration, component performance validation, optical power level adjustment, and research environments requiring high-resolution attenuation control.
Specifications
| Parameter | Value |
|---|---|
| Power Range (dBm) | 30 dBm input (1 W) | Voltage / Current Range: '+5 V output (driver) | Wavelength Range: 1200–1700 nm (standard) |
JDS Uniphase/JDSU HA9/HA1 — Repair & Calibration Service
Aumictech offers NIST-traceable calibration and repair for the JDS Uniphase/JDSU HA9/HA1. Calibration covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.
Common JDS Uniphase/JDSU HA9/HA1 faults: motor and gear mechanism wear (attenuation steps or loss of smooth control), optical port contamination, EEPROM calibration data corruption. Most common faults are motor or gear mechanism wear on variable attenuators causing attenuation steps or loss of smooth control, optical port contamination reducing dynamic range, and EEPROM calibration data corruption on digitally controlled models.
JDS Uniphase/JDSU HA9/HA1 price
This JDS Uniphase/JDSU HA9/HA1 is listed at $2,930. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.
JDS Uniphase/JDSU HA9/HA1 calibration cost
NIST-traceable calibration for the JDS Uniphase/JDSU HA9/HA1 is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. Turnaround is 1 to 3 business days after receipt. See full calibration details →