Anritsu MN924A Optical Attenuator

The Anritsu MN924A is a precision optical attenuator designed for accurate and reliable signal control in fiber optic communication systems. It features a wide attenuation range, low insertion loss, and excellent repeatability, making it suitable for testing and measurement applications. This attenuator supports various fiber types and offers precise adjustment capabilities for optimizing signal levels.

Model
Anritsu MN924A
Calibration
Available on request
Price
$1,200
Status
In stock
RFQ ref
MN924A
View cart
Prefer to talk? Call +1 (484) 841-9341

This unit

The Anritsu MN924A is a precision optical attenuator designed for single-mode fiber applications requiring accurate signal level management across standard telecommunications wavelengths. It combines step and continuous attenuation control in a compact benchtop form factor, delivering up to 65 dB of attenuation with minimal insertion loss and excellent return loss performance.

Technical Specifications

Model: Anritsu MN924A

Fiber Type: Single-mode optical fiber

Wavelengths: 1.3 µm and 1.55 µm (1300 nm / 1550 nm)

Maximum Attenuation: 65 dB

Attenuation Control: Step attenuator (10 dB increments: 0, 10, 20, 30, 40, 50, 60 dB) + continuously variable attenuator (0 to 15 dB in 0.5 dB steps)

Step Attenuation Accuracy: ±1.0 dB (10–20 dB), ±2.0 dB (30–50 dB), ±3.0 dB (≥50 dB)

Variable Attenuation Accuracy: ±1.0 dB (0–10 dB), ±2.0 dB (0–15 dB)

Insertion Loss: ≤3.5 dB

Return Loss: ≥10 dB (FC type connector)

Input-to-Output Crosstalk: ≥70 dB (minimum position)

Optical Connector: FC type

Dimensions: 88 mm (H) × 120 mm (W) × 172 mm (D)

Weight: <2.5 kg

Key Features

Dual attenuation architecture provides coarse (step) and fine (continuous) control for flexible signal conditioning

Low insertion loss maintains optical signal quality across the C and L bands

High crosstalk isolation ensures clean, undegraded optical paths

FC connectors with standard configuration; Super PC-type connectors available on MN9605A variant for enhanced return loss performance

Typical Applications

Fiber optic communication system characterization and testing

Optical power level calibration and verification

Signal attenuation in single-mode fiber networks operating at 1.3 µm or 1.55 µm

Compatibility & Integration The MN924A operates exclusively with single-mode optical fibers and is configured for the 1.3 µm and 1.55 µm telecommunications windows.

Specifications

ParameterValue
Wavelength Range1.3 µm and 1.55 µm (1300 nm / 1550 nm)

Anritsu MN924A — Repair & Calibration Service

Aumictech offers NIST-traceable calibration and repair for the Anritsu MN924A. Calibration covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. All calibrations ship with a NIST-traceable certificate. Turnaround is typically 1–3 business days after receipt.

Standard Cal
$450
Pass/fail cert + NIST sticker
Cal with Data
$650
Full NIST cert with measured values

Common Anritsu MN924A faults: motor and gear mechanism wear (attenuation steps or loss of smooth control), optical port contamination, EEPROM calibration data corruption. Most common faults are motor or gear mechanism wear on variable attenuators causing attenuation steps or loss of smooth control, optical port contamination reducing dynamic range, and EEPROM calibration data corruption on digitally controlled models.

Anritsu MN924A price

This Anritsu MN924A is listed at $1,200. Calibration is available separately from $450. Contact us for volume pricing or if you need multiple units.

Anritsu MN924A calibration cost

NIST-traceable calibration for the Anritsu MN924A is $450 (standard — pass/fail certificate and calibration sticker) or $650 (full measured data at every test point, reference standards documented, traceability chain included). Calibration for this instrument covers attenuation accuracy at 5 or more set values across the full range (0 to rated maximum dB), attenuation linearity, wavelength-dependent attenuation at 1310 nm, 1490 nm, and 1550 nm, insertion loss at minimum attenuation (0 dB position), and polarization-dependent loss (PDL) where applicable. Turnaround is 1 to 3 business days after receipt. See full calibration details →